Ottens Jane, Baker Robert A, Sanderson Andrew J, Newland Richard F
Ashford Hospital, Ashford, South Australia.
J Extra Corpor Technol. 2010 Jun;42(2):153-7.
The use of surface modified, biocompatible tubing in cardiopulmonary bypass has been reported to decrease the inflammatory responses caused by blood contact with the non endothelial surface of poly vinyl chloride (PVC) tubing. The combination of advances in biocompatible tubing and increased affordability resulted in a change to our cardiopulmonary bypass circuit, with the Terumo Capiox SX25 oxygenator and Cobe PVC tubing being replaced with a Terumo Capiox SX25RX (with X coating) and Cobe SMARxT tubing. Prior to the introduction of the coated oxygenator, no connection problems had been evident. One unrelated disconnection involving coated tubing was reported in June 2005 to the Australian and New Zealand College of Perfusionists Perfusion Incident Reporting System. At this time we revised all of our set up protocols and the recommended actions from manufacturers. We further report three separate incidents of pump boot disconnection from the venous reservoir outlet of the oxygenator during bypass (that occurred within a 13-month period), and an outline of immediate and prospective evaluation of the probable cause. We propose that SMARxT 3/8" x 3/32" tubing should not be used on the venous outlet connector of Terumo Capiox SX25RX oxygenators. It appears as though the design of the outlet combined with the properties of SMARxT tubing may contribute to the disconnection.
据报道,在体外循环中使用表面改性的生物相容性管道可减少因血液与聚氯乙烯(PVC)管道的非内皮表面接触而引起的炎症反应。生物相容性管道技术的进步与成本的降低,促使我们对体外循环回路进行了更换,用Terumo Capiox SX25RX(带X涂层)和Cobe SMARxT管道取代了Terumo Capiox SX25氧合器和Cobe PVC管道。在引入涂层氧合器之前,未出现明显的连接问题。2005年6月,澳大利亚和新西兰灌注师学院灌注事件报告系统收到一起与涂层管道无关的断开连接报告。此时,我们修订了所有的设置方案以及制造商推荐的操作方法。我们进一步报告了三起在体外循环期间泵靴从氧合器静脉储液器出口断开连接的独立事件(发生在13个月内),并概述了对可能原因的即时和前瞻性评估。我们建议,Terumo Capiox SX25RX氧合器的静脉出口连接器不应使用SMARxT 3/8" x 3/32"管道。似乎出口的设计与SMARxT管道的特性共同导致了断开连接。