To compare the visual performance after cataract surgery with implantation of +3.00 diopter (D) or +4.00 D aspheric multifocal intraocular lenses (IOLs).
Department of Ophthalmology, Maastricht University Medical Center, The Netherlands.
This prospective study compared the results of bilateral cataract surgery with implantation of a +3.00 D AcrySof ReSTOR SN6AD1 IOL or a +4.00 D AcrySof ReSTOR SN6AD3 IOL. The main outcome measures were binocular uncorrected (UDVA) and corrected distance visual acuities, binocular uncorrected (UNVA) and corrected near visual acuities, binocular uncorrected (UIVA) and corrected intermediate visual acuities, preferred working distance, straylight and contrast sensitivity levels, and wavefront aberrometry measurements.
The +3.00 D IOL was implanted in 68 eyes and the +4.00 D IOL, in 46 eyes. The UIVA was significantly better in the +3.00 D IOL group than in the +4.00 D IOL group at 40, 50, 60, and 70 cm. The preferred working distance for near tasks was significantly lower in the +3.00 D IOL group (38.9 cm) than in the +4.00 D IOL group (31.0 cm). The UDVA was better in the +3.00 D IOL group; the UNVA at the preferred working distance was similar in the 2 groups. Contrast sensitivity and intraocular straylight levels were also similar. The mean levels of higher-order and spherical aberrations were lower in the +3.00 D IOL group.
Cataract surgery with the +3.00 D IOL resulted in better intermediate vision than with the +4.00 D model without compromising distance and near visual acuity.
No author has a financial or proprietary interest in any material or method mentioned.
比较白内障术后植入+3.00 屈光度(D)或+4.00 D 非球面多焦点人工晶状体(IOL)的视觉效果。
荷兰马斯特里赫特大学医学中心眼科。
这项前瞻性研究比较了双侧白内障手术植入+3.00 D AcrySof ReSTOR SN6AD1 IOL 或+4.00 D AcrySof ReSTOR SN6AD3 IOL 的结果。主要观察指标为双眼未矫正(UDVA)和矫正远视力、双眼未矫正(UNVA)和矫正近视力、双眼未矫正(UIVA)和矫正中距离视力、最佳工作距离、散射光和对比敏感度水平以及波前像差测量值。
68 只眼植入+3.00 D IOL,46 只眼植入+4.00 D IOL。在 40、50、60 和 70 cm 时,+3.00 D IOL 组的 UIVA 明显优于+4.00 D IOL 组。+3.00 D IOL 组的近距工作距离明显低于+4.00 D IOL 组(38.9 cm)。+3.00 D IOL 组的 UDVA 较好,两组的近距最佳工作距离 UNVA 相似。对比敏感度和眼内散射光水平也相似。+3.00 D IOL 组的高阶像差和球差平均水平较低。
白内障术后植入+3.00 D IOL 可改善中距离视力,而不会影响远距和近距视力。
无作者在提及的任何材料或方法上具有经济或所有权利益。