使用OPD-Scan III像差仪比较两种多焦点人工晶状体的长期偏心和倾斜情况。

Comparison of long-term decentration and tilt in two types of multifocal intraocular lenses with OPD-Scan III aberrometer.

作者信息

He Wenwen, Qiu Xiaodi, Zhang Shaohua, Du Yu, Zhang Yinglei, Lu Yi, Zhu Xiangjia

机构信息

Eye Institute,Eye and Ear, Nose, and Throat Hospital, Fudan University, 83 Fenyang Road, Shanghai, 200031, People's Republic of China.

Key Laboratory of Myopia, Ministry of Health, Shanghai, 200031, People's Republic of China.

出版信息

Eye (Lond). 2018 Jul;32(7):1237-1243. doi: 10.1038/s41433-018-0068-5. Epub 2018 Mar 9.

Abstract

PURPOSE

To compare the long-term decentration and tilt in two multifocal intraocular lenses (MfIOLs) with an OPD-Scan III aberrometer.

METHODS

Eighty cataract patients who underwent uneventful MfIOL implantation (42 with AcrySof® IQ ReSTOR® SN6AD1 and 38 with AMO Tecnis® ZMB00) were enrolled. At 1 year after surgery, a postoperative visual acuity evaluation included the measurement of uncorrected distance visual acuity, corrected distance visual acuity, distance-corrected near visual acuity, and distance-corrected intermediate visual acuity. OPD-Scan III aberrometer was used to collect the decentration, tilt, and high-order aberration (HOA) data. Significance was tested with Student's t test, Mann-Whitney U test, paired t test, and χ test RESULTS: The average intraocular tilt was lower in the SN6AD1 group than in the ZMB00 group, whereas the average decentration of the two groups did not differ significantly. The mean total ocular HOAs, ocular trefoil, total internal HOAs, and spherical aberrations were significantly lower in the SN6AD1 group than in the ZMB00 group. In both the SN6AD1 and ZMB00 groups, the intraocular tilt was directly proportional to the total ocular HOAs, coma, and spherical aberration. However, there was no significant correlation between decentration and any type of HOA in SN6AD1 group, whereas decentration correlated positively with total ocular HOAs, coma, and spherical aberration in ZMB00 group.

CONCLUSION

Both MfIOLs significantly improve the visual acuity of the patients. However, tilt was lower in the SN6AD1 group than in the ZMB00 group 1 year after surgery, which provide fewer HOAs and better visual quality.

摘要

目的

使用OPD-Scan III像差仪比较两种多焦点人工晶状体(MfIOL)的长期偏心和倾斜情况。

方法

纳入80例接受了平稳MfIOL植入手术的白内障患者(42例植入AcrySof® IQ ReSTOR® SN6AD1,38例植入AMO Tecnis® ZMB00)。术后1年,进行术后视力评估时测量未矫正远视力、矫正远视力、距离矫正近视力和距离矫正中间视力。使用OPD-Scan III像差仪收集偏心、倾斜和高阶像差(HOA)数据。采用Student t检验、Mann-Whitney U检验、配对t检验和χ检验进行显著性检验。结果:SN6AD1组的平均眼内倾斜低于ZMB00组,而两组的平均偏心无显著差异。SN6AD1组的平均总眼HOA、眼三叶草像差、总内部HOA和球差均显著低于ZMB00组。在SN6AD1组和ZMB00组中,眼内倾斜均与总眼HOA、彗差和球差成正比。然而,SN6AD1组的偏心与任何类型的HOA均无显著相关性,而ZMB00组的偏心与总眼HOA、彗差和球差呈正相关。

结论

两种MfIOL均能显著提高患者视力。然而,术后1年SN6AD1组的倾斜低于ZMB00组,其高阶像差较少,视觉质量更好。

相似文献

3
Inferior Decentration of Multifocal Intraocular Lenses in Myopic Eyes.近视眼中多焦点人工晶状体的下偏心。
Am J Ophthalmol. 2018 Apr;188:1-8. doi: 10.1016/j.ajo.2018.01.007. Epub 2018 Jan 31.

引用本文的文献

本文引用的文献

3
Comparison of Ocular Aberrations in Two Hydrophobic and Hydrophilic Intraocular Lenses.
Eye Contact Lens. 2015 Sep;41(5):287-90. doi: 10.1097/ICL.0000000000000118.
8
Clinical evaluation of a multifocal aspheric diffractive intraocular lens.多焦点非球面衍射型人工晶状体的临床评估。
Br J Ophthalmol. 2013 Dec;97(12):1560-4. doi: 10.1136/bjophthalmol-2013-304010. Epub 2013 Oct 11.
9

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验