Thornton A F, Ten Haken R K, Gerhardsson A, Correll M
Department of Radiation Oncology, University of Michigan Medical Center, Ann Arbor 48109-0010.
Radiother Oncol. 1991 Apr;20(4):224-8. doi: 10.1016/0167-8140(91)90120-6.
A mask/marker immobilization system for the routine radiation therapy treatment of head and neck disease is described, utilizing a commercially available thermoplastic mesh, indexed and mounted to a rigid frame attached to the therapy couch. Designed to permit CT, MRI, and PET diagnostic scans of the patient to be performed in the simulation and treatment position employing the same mask, the system has been tested in order to demonstrate the reproducibility of immobilization throughout a radical course of irradiation. Three-dimensional analysis of patient position over an 8-week course of daily radiation treatment has been performed for nine patients from digitization of anatomic points identified on orthogonal radiographs. Studies employing weekly simulation indicate that patient treatment position movement can be restricted to 2 mm over the course of treatment. This easily constructed system permits rapid mask formation to be performed on the treatment simulator, resulting in an immobilization device comparable to masks produced with vacuum-forming techniques. Details of motion analysis and central axis CT, MRI, and PET markers are offered.
描述了一种用于头颈部疾病常规放射治疗的面罩/标记物固定系统,该系统利用市售的热塑性网片,通过索引并安装到附在治疗床上的刚性框架上。该系统设计成允许在模拟和治疗位置使用同一面罩对患者进行CT、MRI和PET诊断扫描,并且已经过测试,以证明在整个根治性放射疗程中固定的可重复性。通过对正交射线照片上确定的解剖点进行数字化处理,对9名患者在为期8周的每日放射治疗过程中的患者位置进行了三维分析。每周进行模拟的研究表明,在治疗过程中患者治疗位置的移动可限制在2毫米以内。这种易于构建的系统允许在治疗模拟器上快速形成面罩,从而产生一种与真空成型技术生产的面罩相当的固定装置。提供了运动分析以及中心轴CT、MRI和PET标记物的详细信息。