Department of Radiology, MR Centre- Highfield MR, Vienna, Austria.
Invest Radiol. 2010 Sep;45(9):554-8. doi: 10.1097/RLI.0b013e3181ebd4e3.
PURPOSE/INTRODUCTION: The aim of this study was to determine the T1 relaxivities (r1) of 8 gadolinium (Gd)-based MR contrast agents in human blood plasma at 7 Tesla, compared with 3 Tesla.
Eight commercially available Gd-based MR contrast agents were diluted in human blood plasma to concentrations of 0, 0.25, 0.5, 1, and 2 mmol/L. In vitro measurements were performed at 37 degrees C, on a 7 Tesla and on a 3 Tesla whole-body magnetic resonance imaging scanner. For the determination of T1 relaxation times, Inversion Recovery Sequences with inversion times from 0 to 3500 ms were used. The relaxivities were calculated.
The r1 relaxivities of all agents, diluted in human blood plasma at body temperature, were lower at 7 Tesla than at 3 Tesla. The values at 3 Tesla were comparable to those published earlier. Notably, in some agents, a minor negative correlation of r1 with a concentration of up to 2 mmol/L could be observed. This was most pronounced in the agents with the highest protein-binding capacity.
DISCUSSION/CONCLUSION: At 7 Tesla, the in vitro r1 relaxivities of Gd-based contrast agents in human blood plasma are lower than those at 3 Tesla. This work may serve as a basis for the application of Gd-based MR contrast agents at 7 Tesla. Further studies are required to optimize the contrast agent dose in vivo.
目的/介绍:本研究旨在确定 8 种钆(Gd)基磁共振对比剂在 7T 与 3T 下的人血浆 T1 弛豫率(r1)。
将 8 种市售 Gd 基磁共振对比剂稀释至人血浆浓度为 0、0.25、0.5、1 和 2 mmol/L。在 37°C 下,于 7T 和 3T 全身磁共振成像扫描仪上进行体外测量。使用反转恢复序列,反转时间从 0 到 3500ms,来确定 T1 弛豫时间。计算弛豫率。
所有在人血浆中稀释至体温的试剂的 r1 弛豫率在 7T 下均低于 3T。在 3T 下的值与早期发表的值相当。值得注意的是,在一些试剂中,可以观察到 r1 与浓度之间的轻微负相关,最大可达 2mmol/L。这在结合能力最高的试剂中最为明显。
讨论/结论:在 7T 下,人血浆中 Gd 基磁共振对比剂的体外 r1 弛豫率低于 3T。本工作可为在 7T 下应用 Gd 基磁共振对比剂提供依据。需要进一步研究以优化体内对比剂剂量。