Suppr超能文献

使用柯达爱克发分析仪进行的分析。采用参考方法值的准确性控制及蛋白质浓度的影响。第二部分。底物。

Analyses with the KODAK-Ektachem. Accuracy control using reference method values and the influence of protein concentration. Part II. Substrates.

作者信息

Külpmann W R, Maibaum P, Sonntag O, Schumann G, Siekmann L

机构信息

Institut für Klinische Chemie I, Medizinische Hochschule Hannover.

出版信息

J Clin Chem Clin Biochem. 1990 Nov;28(11):835-43.

PMID:2077096
Abstract

The reliability of the determination of the most common substrates with the Ektachem 700 was evaluated. Accuracy control was performed in various ways including the comparison of results with reference method values. The influence of protein concentration was investigated systematically using sera containing varying amounts of protein obtained by ultracentrifugation. Bilirubin. Deviation from the reference method values was within limits of the guidelines (Dt. Arztebl. 85 (1988) B517-B532). The influence of protein concentration was negligible. Cholesterol. The results agreed well with the reference method values. The method was not influenced by different protein concentrations. Creatinine. The results were in good agreement with reference method values, when the method was calibrated with the primary assigned values of the calibrators. At high protein concentrations, the results were higher than those of the comparative method. Glucose. The mean deviation from the reference method values was 1.0%. There was a clear positive bias at high protein concentrations. Protein. The results from the Ektachem deviated from the reference method values by -6.8%. Total glycerol (triacylglycerols). In the analysis of control sera, the Ektachem values differed greatly (+32.1%) from the reference method values. This difference was not found in a comparative study with native sera. At high protein concentrations the Ektachem results were higher than those of the comparative method. Urea. The results were lower than the method-dependent assigned values (-16.6%). This deviation was not observed in a comparative study with native sera. Interference by protein was not observed. Uric acid. In accuracy control with reference method values, a small bias was observed (+3.3%), which increased at high protein concentrations.

摘要

评估了使用Ektachem 700测定最常见底物的可靠性。通过多种方式进行准确性控制,包括将结果与参考方法值进行比较。使用通过超速离心获得的含有不同蛋白质含量的血清系统地研究了蛋白质浓度的影响。胆红素。与参考方法值的偏差在指南范围内(德国医学周刊85 (1988) B517 - B532)。蛋白质浓度的影响可忽略不计。胆固醇。结果与参考方法值吻合良好。该方法不受不同蛋白质浓度的影响。肌酐。当用校准品的原始指定值对该方法进行校准时,结果与参考方法值吻合良好。在高蛋白浓度下,结果高于比较方法的结果。葡萄糖。与参考方法值的平均偏差为1.0%。在高蛋白浓度下存在明显的正偏差。蛋白质。Ektachem的结果与参考方法值相差 -6.8%。总甘油(三酰甘油)。在对照血清分析中,Ektachem的值与参考方法值差异很大(+32.1%)。在与天然血清的比较研究中未发现这种差异。在高蛋白浓度下,Ektachem的结果高于比较方法的结果。尿素。结果低于方法相关的指定值(-16.6%)。在与天然血清的比较研究中未观察到这种偏差。未观察到蛋白质的干扰。尿酸。在与参考方法值的准确性控制中,观察到一个小的偏差(+3.3%),在高蛋白浓度下偏差增大。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验