Department of Cardiology, Arrhythmologic Centre, Ospedali del Tigullio, Lavagna, Italy.
Europace. 2010 Nov;12(11):1616-22. doi: 10.1093/europace/euq333. Epub 2010 Sep 13.
We thought to evaluate feasibility of continuous non-invasive blood pressure monitoring during procedures of interventional electrophysiology.
We evaluated continuous non-invasive finger blood pressure (BP) monitoring by means of the Nexfin device in 22 patients (mean age 70 ± 24 years), undergoing procedures of interventional electrophysiology, in critical situations of hypotension caused by tachyarrhythmias or by intermittent incremental ventricular temporary pacing till to the maximum tolerated systolic BP fall (mean 61 ± 14 mmHg per patient at a rate of 195 ± 37 bpm). In all patients, Nexfin was able to detect immediately, at the onset of tachyarrythmia, the changes in BP and recorded reliable waveforms. The quality of the signal was arbitrarily classified as excellent in 11 cases, good in 10 cases, and sufficient in 1 case. In basal conditions, calibrations of the signal occurred every 49.2 ± 24.3 s and accounted for 4% of total monitoring time; during tachyarrhythmias their frequency increased to one every 12.7 s and accounted for 19% of total recording duration. A linear correlation for a range of BP values from 41 to 190 mmHg was found between non-invasive and intra-arterial BP among a total of 1055 beats from three patients who underwent simultaneous recordings with both methods (coefficient of correlation of 0.81, P < 0.0001).
In conclusion, continuous non-invasive BP monitoring is feasible in the clinical practise of an interventional electrophysiology laboratory without the need of utilization of an intra-arterial BP line.
我们旨在评估在介入电生理程序期间进行连续无创血压监测的可行性。
我们评估了 22 名(平均年龄 70±24 岁)接受介入电生理程序的患者在心动过速或间歇性递增心室临时起搏引起低血压的危急情况下通过 Nexfin 设备进行连续无创手指血压(BP)监测的可行性,直到最大耐受收缩压下降(平均每位患者每分钟 195±37 次心动过速时收缩压下降 61±14mmHg)。在所有患者中,Nexfin 能够立即检测到心动过速开始时的血压变化,并记录可靠的波形。信号质量被任意分为 11 例优秀、10 例良好和 1 例充足。在基础条件下,信号校准每 49.2±24.3 秒发生一次,占总监测时间的 4%;在心动过速期间,其频率增加到每 12.7 秒一次,占总记录时间的 19%。在同时使用两种方法的 3 名患者的总共 1055 次心跳中,发现无创血压和动脉内血压之间在 41 至 190mmHg 的血压范围内存在线性相关(相关系数为 0.81,P<0.0001)。
总之,在介入电生理实验室的临床实践中,连续无创血压监测是可行的,无需使用动脉内血压线。