Biosens Bioelectron. 2010 Nov 15;26(3):1109-13. doi: 10.1016/j.bios.2010.08.021. Epub 2010 Aug 20.
A CdSe/ZnS quantum dot (QD) based electrochemical immunoassay of phosphorylated bovine serum albumin (BSA-OP) as a protein biomarker is presented. The QDs were used as labels for amplifying electrochemical signals and were conjugated with a secondary anti-phosphoserine antibody in a heterogeneous sandwich immunoassay. In this assay, the model phosphorylated protein BSA-OP was added to the primary BSA antibody coated polystyrene microwells, and then the QD labeled anti-phosphoserine antibody was added for completing immunorecognition. Finally, the bound QD was dissolved in an acid-dissolution step and was detected by electrochemical stripping analysis. The measured current responses were proportional to the concentration of BSA-OP. Under optimal conditions, the voltammetric response was linear over the range of 0.5-500 ngmL(-1) of BSA-OP, with a detection limit of 0.5 ngmL(-1). It also shows good reproducibility with a relative standard deviation of 8.6%. This QD-based electrochemical immunoassay offers great promise for simple and cost-effective analysis of protein biomarkers.
基于 CdSe/ZnS 量子点的电化学免疫分析检测磷酸化牛血清白蛋白作为一种蛋白质生物标志物。量子点被用作电化学信号放大的标签,并与一种抗磷酸丝氨酸的二级抗体在异质三明治免疫分析中结合。在该分析中,模型磷酸化蛋白 BSA-OP 被加入到涂有原发性 BSA 抗体的聚苯乙烯微孔中,然后加入量子点标记的抗磷酸丝氨酸抗体以完成免疫识别。最后,结合的量子点在酸溶解步骤中溶解,并通过电化学剥离分析进行检测。测量的电流响应与 BSA-OP 的浓度成正比。在最佳条件下,伏安响应在 0.5-500ngmL(-1)BSA-OP 的范围内呈线性,检测限为 0.5ngmL(-1)。它还表现出良好的重现性,相对标准偏差为 8.6%。这种基于量子点的电化学免疫分析为简单、经济有效地分析蛋白质生物标志物提供了很大的前景。