Platelet aggregation monitoring with a newly developed quartz crystal microbalance system as an alternative to optical platelet aggregometry.

机构信息

Children's University Hospital, Div. Congenital & Pediatric Cardiac Surgery, Clinical Research Laboratory, Tuebingen University, Calwerstr. 7/1, 72076 Tuebingen, Germany.

出版信息

Analyst. 2010 Nov;135(11):2930-8. doi: 10.1039/c0an00474j. Epub 2010 Sep 29.

Abstract

The objective of this study was to establish a new test system for the monitoring of platelet aggregation during extracorporeal circulation (ECC) procedures. Even though extensive progress has been made in improving the haemocompatibility of extracorporeal circulation devices, activation of blood coagulation, blood platelets and inflammatory responses are still undesired outcomes of cardiopulmonary bypass. This study deals with an approach towards a platelet aggregation measuring system using a newly developed quartz crystal microbalance (QCM) system. Since QCM is a rarely used technique in the field of blood analytics, the challenge was to transfer the well established methods of aggregometry to the new test system. In a QCM system, either bare gold or fibrinogen-coated sensors were incubated with ADP or arachidonic acid (AA) stimulated platelet rich plasma. For negative controls the GPIIb/IIIa inhibitory antibody abciximab (Reopro®) was used as an inhibitor of platelet aggregation. During incubation, the frequency shifts of the sensors were recorded. The results gained from the QCM system were compared to results gained by optical platelet aggregometry (born aggregometry). For additional visualization of platelet adhesion to the sensor surfaces, fluorescent microscopy and scanning electron microscopy were used. The QCM sensor was able to detect platelet aggregation in both uncoated and fibrinogen coated sensors. The measuring curves of aggregation measurements and controls were clearly distinguishable from each other in terms of frequency shifts and kinetics. For aggregation measurements and inhibited controls the therapeutic diagnosis of platelet function is identical between aggregometer and QCM data. In future, QCM based measuring devices may become an alternative to established point of care methods for rapid bedside testing of platelet aggregation.

摘要

本研究旨在建立一种新的体外循环(ECC)过程中血小板聚集监测测试系统。尽管在改善体外循环设备的血液相容性方面已经取得了广泛的进展,但凝血、血小板激活和炎症反应仍然是体外循环的不良后果。本研究涉及一种使用新开发的石英晶体微天平(QCM)系统测量血小板聚集的方法。由于 QCM 在血液分析领域很少使用,因此挑战在于将成熟的聚集度测定方法应用于新的测试系统。在 QCM 系统中,裸金或纤维蛋白原涂层传感器与 ADP 或花生四烯酸(AA)刺激的富含血小板血浆孵育。作为血小板聚集抑制剂,GPIIb/IIIa 抑制性抗体 abciximab(Reopro®)用于阴性对照。在孵育过程中,记录传感器的频率变化。从 QCM 系统获得的结果与光学血小板聚集度测定(比浊法)获得的结果进行比较。为了进一步观察血小板在传感器表面的粘附,使用荧光显微镜和扫描电子显微镜。QCM 传感器能够在未涂层和纤维蛋白原涂层传感器中检测到血小板聚集。在频率变化和动力学方面,聚集度测量和对照的测量曲线明显彼此区分开来。对于聚集度测量和抑制对照,血小板功能的治疗诊断在聚集仪和 QCM 数据之间是相同的。在未来,基于 QCM 的测量设备可能成为替代现有的床边即时检测血小板聚集的方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验