Nat Commun. 2022 Feb 11;13(1):831. doi: 10.1038/s41467-022-28499-y.
Frequent prothrombin time (PT) and international normalized ratio (INR) testing is critical for millions of people on lifelong anticoagulation with warfarin. Currently, testing is performed in hospital laboratories or with expensive point-of-care devices limiting the ability to test frequently and affordably. We report a proof-of-concept PT/INR testing system that uses the vibration motor and camera on smartphones to track micro-mechanical movements of a copper particle. The smartphone system computed the PT/INR with inter-class correlation coefficients of 0.963 and 0.966, compared to a clinical-grade coagulation analyzer for 140 plasma samples and demonstrated similar results for 80 whole blood samples using a single drop of blood (10 μl). When tested with 79 blood samples with coagulopathic conditions, the smartphone system demonstrated a correlation of 0.974 for both PT/INR. Given the ubiquity of smartphones in the global setting, this proof-of-concept technology may provide affordable and effective PT and INR testing in low-resource environments.
对于数百万需要终身接受华法林抗凝治疗的人来说,频繁进行凝血酶原时间(PT)和国际标准化比值(INR)检测至关重要。目前,检测是在医院实验室进行的,或者使用昂贵的即时检测设备进行,这限制了频繁且经济地检测的能力。我们报告了一种使用智能手机上的振动电机和摄像头来跟踪铜颗粒微机械运动的概念验证 PT/INR 检测系统。智能手机系统通过对 140 个血浆样本与临床级凝血分析仪进行对比,计算出 PT/INR,得出的组内相关系数为 0.963 和 0.966,对于 80 个全血样本,仅用一滴血(10μl)即可获得相似的结果。在对 79 份有凝血功能障碍的血液样本进行测试时,智能手机系统对 PT/INR 的相关性为 0.974。鉴于智能手机在全球范围内的普及,这项概念验证技术可能为资源匮乏环境提供负担得起且有效的 PT 和 INR 检测。