Second Sight Medical Products, Sylmar, CA 91342, USA.
Br J Ophthalmol. 2011 Apr;95(4):539-43. doi: 10.1136/bjo.2010.179622. Epub 2010 Sep 29.
BACKGROUND/AIMS: To determine to what extent subjects implanted with the Argus II retinal prosthesis can improve performance compared with residual native vision in a spatial-motor task.
High-contrast square stimuli (5.85 cm sides) were displayed in random locations on a 19″ (48.3 cm) touch screen monitor located 12″ (30.5 cm) in front of the subject. Subjects were instructed to locate and touch the square centre with the system on and then off (40 trials each). The coordinates of the square centre and location touched were recorded.
Ninety-six percent (26/27) of subjects showed a significant improvement in accuracy and 93% (25/27) show a significant improvement in repeatability with the system on compared with off (p<0.05, Student t test). A group of five subjects that had both accuracy and repeatability values <250 pixels (7.4 cm) with the system off (ie, using only their residual vision) was significantly more accurate and repeatable than the remainder of the cohort (p<0.01). Of this group, four subjects showed a significant improvement in both accuracy and repeatability with the system on.
In a study on the largest cohort of visual prosthesis recipients to date, we found that artificial vision augments information from existing vision in a spatial-motor task. Clinical trials registry no NCT00407602.
背景/目的:确定植入 Argus II 视网膜假体的受试者在空间运动任务中与残留的自然视力相比,能在多大程度上提高表现。
高对比度的正方形刺激物(5.85 厘米边长)随机显示在位于受试者前方 12 英寸(30.5 厘米)的 19 英寸(48.3 厘米)触摸屏显示器的任意位置。受试者被指示用系统开启和关闭时分别找到并触摸正方形中心(各 40 次)。记录正方形中心和触摸位置的坐标。
96%(26/27)的受试者在系统开启时的准确性显著提高,93%(25/27)的受试者在系统开启时的重复性显著提高(p<0.05,学生 t 检验)。一组五名受试者在系统关闭时(即仅使用其残留视力)的准确性和重复性值均<250 像素(7.4 厘米),其准确性和重复性显著高于其余受试者(p<0.01)。这组中有四名受试者在系统开启时准确性和重复性均有显著提高。
在迄今为止对最大视觉假体受者队列的研究中,我们发现人工视觉在空间运动任务中增强了现有视觉的信息。临床试验注册号 NCT00407602。