Department of Epidemiology and the Welch Center for Prevention, Epidemiology and Clinical Research, Johns Hopkins Bloomberg School of Public Health, Baltimore, Maryland 21287, USA.
J Diabetes. 2010 Jun;2(2):118-24. doi: 10.1111/j.1753-0407.2010.00070.x. Epub 2010 Mar 15.
The aims of the present study were to demonstrate the reliability of HbA1c measurements during two time periods and to compare these measurements with HbA1c distribution in the general US population.
HbA1c was measured in 14,069 whole blood samples in the Atherosclerosis Risk in Communities (ARIC) study using different HPLC instruments across two time periods, namely 2003-2004 and 2007-2008. At the time of measurement, samples had been in storage at -70°C for up to 18 years. To assess differences in values, HbA1c measurements were repeated in 383 samples at both periods. Indirect comparisons were made by comparing our measurements against those from a nationally representative study.
The coefficients of variation for quality control samples were 1.8% (n = 89) in 2003-2004 and 1.4% (n = 259) in 2007-2008. The correlation between measurements at the two time points was high (r = 0.99), but with a slight bias: 0.29% points higher in 2007-2008 vs 2003-2004 (n = 383; P < 0.0001). The comparison yielded the following Deming regression equation: y((2007-2008)) = 0.073 + 1.034x((2003-2004)) . After alignment using this equation, the distribution of HbA1c in the ARIC study was similar to that in the national study using fresh samples.
Measurements of HbA1c from samples stored for up to 18 years are highly reliable when using state-of-the-art HPLC instruments, but with some bias introduced over time. The HbA1c data now available in the ARIC study should be invaluable for investigations into the clinical utility of HbA1c as a diagnostic test for diabetes.
本研究旨在证明 HbA1c 在两个时间段内的测量可靠性,并将这些测量结果与美国一般人群中 HbA1c 的分布进行比较。
在动脉粥样硬化风险社区(ARIC)研究中,使用不同的 HPLC 仪器在两个时间段内(即 2003-2004 年和 2007-2008 年)测量了 14069 份全血样本中的 HbA1c。在测量时,样本已在-70°C 下储存长达 18 年。为了评估值之间的差异,在两个时间段内对 383 个样本重复测量了 HbA1c。通过将我们的测量结果与全国代表性研究进行间接比较来进行比较。
质量控制样本的变异系数在 2003-2004 年为 1.8%(n=89),在 2007-2008 年为 1.4%(n=259)。两个时间点之间的测量相关性很高(r=0.99),但存在轻微偏差:2007-2008 年比 2003-2004 年高 0.29%(n=383;P<0.0001)。比较得出以下 Deming 回归方程:y((2007-2008))=0.073+1.034x((2003-2004))。使用该方程进行调整后,ARIC 研究中 HbA1c 的分布与使用新鲜样本的全国研究相似。
使用最先进的 HPLC 仪器,储存长达 18 年的 HbA1c 样本的测量结果高度可靠,但随着时间的推移会出现一些偏差。ARIC 研究中现在可用的 HbA1c 数据对于研究 HbA1c 作为糖尿病诊断测试的临床实用性应该是非常有价值的。