Department of Electrical & Electronic Engineering, The University of Melbourne, Melbourne 3010, Australia.
Med Eng Phys. 2011 Mar;33(2):204-9. doi: 10.1016/j.medengphy.2010.09.020. Epub 2010 Oct 25.
We investigate whether pulse rate variability (PRV) extracted from finger photo-plethysmography (Pleth) waveforms can be the substitute of heart rate variability (HRV) from RR intervals of ECG signals during obstructive sleep apnea (OSA). Simultaneous measurements (ECG and Pleth) were taken from 29 healthy subjects during normal (undisturbed sleep) breathing and 22 patients with OSA during OSA events. Highly significant (p<0.01) correlations (1.0>r>0.95) were found between heart rate (HR) and pulse rate (PR). Bland-Altman plot of HR and PR shows good agreement (<5% difference). Comparison of 2 min recording epochs demonstrated significant differences (p<0.01) in time, frequency domains and complexity analysis, between normal and OSA events using PRV as well as HRV measures. Results suggest that both HRV and PRV indices could be used to distinguish OSA events from normal breathing during sleep. However, several variability measures (SDNN, RMSSD, HF power, LF/HF and sample entropy) of PR and HR were found to be significantly (p<0.01) different during OSA events. Therefore, we conclude that PRV provides accurate inter-pulse variability to measure heart rate variability under normal breathing in sleep but does not precisely reflect HRV in sleep disordered breathing.
我们研究了从手指光体积描记( pleth )波形中提取的脉搏率变异性( PRV )是否可以替代心电图信号 RR 间隔中的心率变异性( HRV )在阻塞性睡眠呼吸暂停( OSA )期间。从 29 名健康受试者在正常(无干扰的睡眠)呼吸期间和 22 名 OSA 患者在 OSA 事件期间同时进行了( ECG 和 pleth )测量。心率( HR )和脉搏率( PR )之间存在高度显著的相关性( 1.0>r>0.95 )( p<0.01 )。 HR 和 PR 的 Bland-Altman 图显示出良好的一致性( <5%差异)。使用 PRV 以及 HRV 指标比较 2 分钟记录时段,发现正常和 OSA 事件之间在时域、频域和复杂性分析方面存在显著差异( p<0.01 )。结果表明,HRV 和 PRV 指数都可用于区分睡眠期间 OSA 事件和正常呼吸。然而,在 OSA 事件期间, PR 和 HR 的几个变异性测量( SDNN 、 RMSSD 、 HF 功率、 LF/HF 和样本熵)被发现存在显著差异( p<0.01 )。因此,我们得出结论,PRV 提供了准确的脉冲间变异性,可在睡眠期间正常呼吸下测量心率变异性,但不能准确反映睡眠呼吸障碍期间的 HRV 。