Suppr超能文献

信号强度对 Stratus 和 Cirrus 光学相干断层扫描仪测量视网膜神经纤维层厚度及其彩色编码分类的一致性的影响。

Effect of signal strength on agreements for retinal nerve fiber layer thickness measurement and its color code classification between Stratus and Cirrus optical coherence tomography.

机构信息

Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea.

出版信息

J Glaucoma. 2011 Aug;20(6):371-6. doi: 10.1097/IJG.0b013e3181efb3a7.

Abstract

PURPOSE

To assess the effect of signal strength (SS) on concordance rate or agreement on retinal nerve fiber layer (RNFL) thickness measurement (measurement agreement) and its color code classification (classification agreement) between Stratus optical coherence tomography (OCT) and Cirrus OCT.

PATIENTS AND METHODS

From 300 participants who underwent both Stratus and Cirrus OCT scans at the same visit (SS ≥ 6), intraclass correlation and weighted κ coefficients were calculated as indicators for RNFL measurement agreement and classification agreement between 2 OCTs.

RESULTS

In inferior and superior quadrants, agreements were excellent (intraclass correlation) or good (κ) and neither the measurement nor the classification agreement was statistically different between participants grouped by SS difference between Stratus and Cirrus OCTs (inter-OCT SS difference). In nasal and temporal quadrants, agreements were low and there were groups with inter-OCT SS difference whose classification agreement between 2 OCTs was not better than the chance (κ=0).

CONCLUSIONS

When comparing RNFL thickness estimated by Cirrus OCT with that measured by Stratus OCT in temporal and nasal quadrants, special attention should be given, as the agreements or concordance rate was low, suggesting that neither the measured value nor its probability algorithm classification would be interchangeable. In inferior and superior quadrants, agreements were good or excellent and not affected by SS difference as long as SS ≥ 6.

摘要

目的

评估信号强度(SS)对视网膜神经纤维层(RNFL)厚度测量(测量一致性)及其彩色编码分类(分类一致性)在 Stratus 光学相干断层扫描(OCT)和 Cirrus OCT 之间的一致性的影响。

方法

从 300 名在同一次就诊时同时接受 Stratus 和 Cirrus OCT 扫描的参与者(SS≥6)中,计算了内类相关和加权κ系数,作为 2 种 OCT 之间 RNFL 测量一致性和分类一致性的指标。

结果

在下方和上方象限,一致性极好(内类相关)或良好(κ),并且根据 Stratus 和 Cirrus OCT 之间 SS 差异(眼内 SS 差异)对参与者进行分组时,测量一致性和分类一致性均无统计学差异。在鼻侧和颞侧象限,一致性较低,存在一些组的眼内 SS 差异,其 2 种 OCT 之间的分类一致性并不优于偶然(κ=0)。

结论

在比较 Cirrus OCT 估计的 RNFL 厚度与 Stratus OCT 在颞侧和鼻侧象限测量的厚度时,应特别注意,因为一致性或一致性率较低,表明测量值及其概率算法分类均不可互换。在下方和上方象限,只要 SS≥6,一致性良好或极好,不受 SS 差异的影响。