比较时域和频域光学相干断层扫描测量白内障和瞳孔大小对视网膜神经纤维层厚度的影响。

Comparison of the influence of cataract and pupil size on retinal nerve fibre layer thickness measurements with time-domain and spectral-domain optical coherence tomography.

机构信息

Department of Ophthalmology Clinical Research Unit, Tan Tock Seng Hospital, Singapore.

出版信息

Clin Exp Ophthalmol. 2011 Apr;39(3):215-21. doi: 10.1111/j.1442-9071.2010.02460.x. Epub 2011 Jan 14.

Abstract

BACKGROUND

To investigate and compare the effect of cataract and pupil size on retinal nerve fibre layer (RNFL) thickness measurements using spectral-domain optical coherence tomography (Cirrus OCT) and time-domain OCT (Stratus OCT).

DESIGN

Prospective, hospital-based study.

PARTICIPANTS

Twenty-five eyes from 25 normal subjects undergoing cataract surgery.

METHODS

Three retinal nerve fibre layer (RNFL) thickness measurements were taken before and after dilation, preoperatively and postoperatively, using Cirrus 200×200 Optic Disc Scan and Stratus Fast RNFL Scan.

MAIN OUTCOME MEASURES

Linear regression, intraclass correlation coefficient (ICC) and coefficient of variation analysis.

RESULTS

Cataract removal caused significant increase in RNFL measurements in both modalities (Cirrus P<0.02; Stratus P<0.001). There was no significant difference in the increase in measurements between the two machines. Pupil dilation had variable and non-statistically significant effect in both (P>0.05). ICC showed excellent reproducibility with Cirrus OCT after mydriasis, preoperatively (ICC=0.78-0.90) and postoperatively (ICC=0.90-0.97), but poor reproducibility before mydriasis (P<0.75). Stratus OCT achieved excellent reproducibility after cataract removal both before (ICC=0.86-0.96) and after mydriasis (ICC=0.92-0.95), but poor reproducibility before cataract surgery (P<0.75).

CONCLUSIONS

Cataracts, not pupil size, cause significant underestimation of RNFL measurements in both Cirrus and Stratus OCT. The extent of influence exerted does not appear different between the two instruments. Reproducibility of each machine appears to be affected differently. Mydriasis is required to achieve excellent reproducibility with Cirrus OCT, and media clarity is required with Stratus OCT.

摘要

背景

使用频域光相干断层扫描仪(Cirrus OCT)和时域光相干断层扫描仪(Stratus OCT)来比较白内障和瞳孔大小对视网膜神经纤维层(RNFL)厚度测量的影响。

设计

前瞻性、基于医院的研究。

参与者

25 名接受白内障手术的正常受试者的 25 只眼。

方法

使用 Cirrus 200×200 视盘扫描和 Stratus Fast RNFL 扫描,在散瞳前后、术前和术后进行 3 次视网膜神经纤维层(RNFL)厚度测量。

主要观察指标

线性回归、组内相关系数(ICC)和变异系数分析。

结果

白内障切除后两种模式下的 RNFL 测量值均显著增加(Cirrus P<0.02;Stratus P<0.001)。两种仪器的测量值增加无显著差异。瞳孔扩张在两种情况下均有不同程度的、无统计学意义的影响(P>0.05)。散瞳后 Cirrus OCT 的 ICC 显示出极好的可重复性,术前(ICC=0.78-0.90)和术后(ICC=0.90-0.97),但散瞳前的可重复性较差(P<0.75)。白内障切除后,Stratus OCT 在白内障切除前后均具有极好的可重复性(ICC=0.86-0.96 和 ICC=0.92-0.95),但白内障手术前的可重复性较差(P<0.75)。

结论

白内障而非瞳孔大小导致 Cirrus 和 Stratus OCT 中 RNFL 测量值的显著低估。两种仪器的影响程度似乎没有差异。每台机器的重复性似乎受到不同的影响。Cirrus OCT 需要散瞳才能获得极好的重复性,Stratus OCT 需要清晰的介质。