Kirby E J, Buchalter J S, Kastenbaum D M, Kenan S, Kummer F J, Lewis M M
Department of Orthopaedics, Hospital for Joint Diseases Orthopaedic Institute, New York, NY 10003.
Clin Orthop Relat Res. 1990 Apr(253):231-9.
The CO2 laser was evaluated to determine its effectiveness for marginal cauterization of giant-cell tumors after resectional curettage. Average power densities of 11-53 W/cm2 for corresponding application times of 120 and 60 seconds were found to produce 30 degrees temperature rises in fresh human femoral heads. This temperature was adequate to achieve thermal necrosis of the tumor margins to a depth of 5 mm. CO2 laser resection was found to be accurate and reproducible. No damage to underlying hyaline cartilage and soft tissues was evident histologically at one and four weeks after experimental local laser cauterization of distal femoral condyles in rabbits.
对二氧化碳激光进行评估,以确定其在刮除术后对巨细胞瘤进行边缘烧灼的有效性。发现在120秒和60秒的相应应用时间下,平均功率密度为11 - 53瓦/平方厘米时,新鲜人体股骨头的温度会升高30度。该温度足以使肿瘤边缘达到5毫米深度的热坏死。发现二氧化碳激光切除准确且可重复。在对兔股骨远端髁进行实验性局部激光烧灼后1周和4周,组织学检查未发现对下方透明软骨和软组织有明显损伤。