Graduate Program in Odontology, Brazil.
Dent Mater. 2011 Apr;27(4):e60-4. doi: 10.1016/j.dental.2010.11.010. Epub 2010 Dec 15.
The purpose of the present research was to demonstrate images of failures into the structure of pit and fissure sealants using optical coherence tomography (OCT).
Five human third molars were selected and the sealant ALPHA SEAL LIGHT/DFL was applied according to the manufacturer instructions. For evaluation of the structure of pit and fissure sealants, twenty OCT images of each tooth were performed before and after application of sealant. A total of 200 images were evaluated. A home built spectral OCT system used was operating in spectral domain (SD-OCT), at 840 nm and a measured spatial resolution of 10 μm. The system is based on the Michelson interferometer set-up and is controlled by the software OCT 800--Complex Square/LabView, that collects data and generates the image. The occlusal surfaces were scanned in a bucco-lingual direction and tomographic images parallel to long axis of tooth were obtained. After the achievement of the images by OCT, the crown of each tooth was sectioned in a occlusogingival direction (buccolingually). Images from each section were obtained and evaluated by an optical microscope. OCT and microscopic images were compared.
Representative images showed that OCT image provides the insight into sealant material. It was possible to clearly identify the internal structure of sealant and the subjacent enamel. Sealant and enamel are very well distinguished, and failures at the interior and at surface are well detected.
OCT could generate images of the features of fissures, bubbles and failures in the adaptation of sealants, opening up possibilities in the future to monitoring sealant application and retention in short and long term.
本研究的目的是使用光学相干断层扫描(OCT)显示窝沟封闭剂结构中的失败图像。
选择五个人类第三磨牙,并按照制造商的说明应用 ALPHA SEAL LIGHT/DFL 密封剂。为了评估窝沟封闭剂的结构,在应用密封剂之前和之后对每颗牙齿进行了二十次 OCT 成像。总共评估了 200 个图像。使用的自建光谱 OCT 系统在光谱域(SD-OCT)中运行,工作波长为 840nm,测量的空间分辨率为 10μm。该系统基于迈克尔逊干涉仪设置,并由 OCT 800--Complex Square/LabView 软件控制,该软件可采集数据并生成图像。在颊舌方向上扫描咬合面,并获得与牙齿长轴平行的层析成像。通过 OCT 获得图像后,将每个牙齿的牙冠沿近远中方向(颊舌向)进行切片。从每个切片中获得图像并通过光学显微镜进行评估。将 OCT 和显微镜图像进行比较。
代表性图像表明,OCT 图像提供了对密封剂材料的深入了解。可以清楚地识别密封剂的内部结构和下方的釉质。密封剂和釉质区分得非常好,并且可以很好地检测到内部和表面的失败。
OCT 可以生成窝沟、气泡和封闭剂适应性失败的特征图像,为未来监测短期和长期的密封剂应用和保留提供了可能性。