Performance Characteristics of BGO Detectors for a Low Cost Preclinical PET Scanner.

作者信息

Zhang H, Vu N T, Bao Q, Silverman R W, Berry-Pusey B N, Douraghy A, Williams D A, Rannou F R, Stout D B, Chatziioannou A F

机构信息

UCLA David Geffen School of Medicine, Crump Institute for Molecular Imaging, Los Angeles, CA 90095 USA, on leave from the Department of Biomedical Engineering, Tsinghua University, Beijing 100084, China.

出版信息

IEEE Trans Nucl Sci. 2010 Jun 1;57(3):1038-1044. doi: 10.1109/TNS.2010.2046753.

Abstract

PETbox is a low-cost benchtop PET scanner dedicated to high throughput preclinical imaging that is currently under development at our institute. This paper presents the design and characterization of the detectors that are used in the PETbox system. In this work, bismuth germanate scintillator was used for the detector, taking advantage of its high stopping power, high photoelectric event fraction, lack of intrinsic background radiation and low cost. The detector block was segmented into a pixelated array consisting of 20 × 44 elements, with a crystal pitch of 2.2 mm and a crystal cross section of 2 mm × 2 mm. The effective area of the array was 44 mm × 96.8 mm. The array was coupled to two Hamamatsu H8500 position sensitive photomultiplier tubes, forming a flat-panel type detector head with a sensitive area large enough to cover the whole body of a typical laboratory mouse. Two such detector heads were constructed and their performance was characterized. For one detector head, the energy resolution ranged from 16.1% to 38.5% full width at half maximum (FWHM), with a mean of 20.1%; for the other detector head, the energy resolution ranged from 15.5% to 42.7% FWHM, with a mean of 19.6%. The intrinsic spatial resolution was measured to range from 1.55 mm to 2.39 mm FWHM along the detector short axis and from 1.48 mm to 2.33 mm FWHM along the detector long axis, with an average of 1.78 mm. Coincidence timing resolution for the detector pair was measured to be 4.1 ns FWHM. These measurement results show that the detectors are suitable for our specific application.

摘要

PETbox是一种低成本的台式PET扫描仪,专门用于高通量临床前成像,目前正在我们研究所进行开发。本文介绍了PETbox系统中使用的探测器的设计和特性。在这项工作中,锗酸铋闪烁体被用于探测器,利用了其高阻止本领、高光电事件分数、无固有本底辐射以及低成本的特点。探测器模块被分割成一个由20×44个单元组成的像素化阵列,晶体间距为2.2毫米,晶体横截面为2毫米×2毫米。阵列的有效面积为44毫米×96.8毫米。该阵列与两个滨松H8500位置灵敏光电倍增管耦合,形成一个平板型探测器头,其灵敏面积大到足以覆盖典型实验小鼠的全身。构建了两个这样的探测器头并对其性能进行了表征。对于一个探测器头,能量分辨率在半高宽(FWHM)处为16.1%至38.5%,平均值为20.1%;对于另一个探测器头,能量分辨率在FWHM处为15.5%至42.7%,平均值为19.6%。本征空间分辨率沿探测器短轴测量范围为1.55毫米至2.39毫米FWHM,沿探测器长轴测量范围为1.48毫米至2.33毫米FWHM,平均值为1.78毫米。探测器对的符合时间分辨率测量为4.1纳秒FWHM。这些测量结果表明这些探测器适用于我们的特定应用。

相似文献

引用本文的文献

3
Performance of Dual-Ended Readout PET Detectors Based on BGO Arrays and BaSO Reflector.基于BGO阵列和硫酸钡反射器的双端读出PET探测器的性能
IEEE Trans Radiat Plasma Med Sci. 2022 May;6(5):522-528. doi: 10.1109/trpms.2021.3096534. Epub 2021 Jul 12.
10
Evaluation of the Genisys4, a bench-top preclinical PET scanner.Genisys4 临床前 PET 扫描仪的评估
J Nucl Med. 2013 Jul;54(7):1162-7. doi: 10.2967/jnumed.112.114926. Epub 2013 Apr 29.

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验