Procera®定制基台与三种种植系统之间的适配精度比较。

Comparison of fit accuracy between Procera® custom abutments and three implant systems.

作者信息

de Morais Alves da Cunha Tiago, de Araújo Roberto Paulo Correia, da Rocha Paulo Vicente Barbosa, Amoedo Rosa Maria Pazos

机构信息

Department of Prosthodontics, School of Dentistry, UNIME, Bahia, Brazil.

出版信息

Clin Implant Dent Relat Res. 2012 Dec;14(6):890-5. doi: 10.1111/j.1708-8208.2010.00323.x. Epub 2010 Dec 22.

Abstract

BACKGROUND

Although increase of misfit has been reported when associating implant and abutment from different manufacturers, Procera custom abutment has been universally used in clinical practice.

PURPOSE

The purpose of this investigation was to compare the vertical gap of zirconia Procera® abutment associated with implants from the same manufacturer (Procera manufacturer) and two other implant systems.

MATERIALS AND METHODS

Twenty-four zirconia Procera abutments were produced using computer-assisted design and manufacture and paired with (a) eight MK III, RP 4.1 × 10 mm implants (Nobel Biocare™, Göteborg, Sweden) - GNB group (Nobel Biocare group); (b) eight Try on, 4.1 × 10 mm implants (Sistema de Implantes, São Paulo, Brazil) - ES group (SIN experimental group) ; and (c) eight Master screw, 4.1 × 10 mm implants (Conexão® Sistema de Prótese, São Paulo, Brazil) - EC group (Conexão experimental group). A comparison of the vertical misfit at the implant-abutment interface was taken at six measuring sites on each sample using scanning electron microscopy with a magnification of 408×. One-way analysis of variance was used to test for differences, and Tukey's test was used for pair-wise comparison of groups (α = 0.05).

RESULTS

Significant differences relative to average misfit were found when Procera abutments were associated with other implant manufacturers. The ES group and EC group did not differ significantly, but both demonstrated significantly larger average misfit than the GNB group (p = .001). The average misfit was 5.7 µm ± 0.39, 9.53 µm ± 0.52, and 10.62 µm ± 2.16, respectively, for groups GNB, ES, and EC.

CONCLUSION

The association of Procera zirconia abutment with other implant systems different from its manufacturer demonstrated significant alteration of vertical misfit at implant-abutment interface.

摘要

背景

尽管有报道称,当将不同制造商生产的种植体和基台连接在一起时,不匹配度会增加,但Procera定制基台已在临床实践中广泛使用。

目的

本研究的目的是比较氧化锆Procera®基台与同一制造商(Procera制造商)生产的种植体以及其他两种种植系统连接时的垂直间隙。

材料与方法

使用计算机辅助设计和制造技术制作了24个氧化锆Procera基台,并分别与以下种植体配对:(a) 8个MK III、RP 4.1×10 mm种植体(诺贝尔生物科技公司,瑞典哥德堡)——GNB组(诺贝尔生物科技组);(b) 8个试戴型、4.1×10 mm种植体(巴西圣保罗植入系统公司)——ES组(SIN实验组);以及(c) 8个主螺杆、4.1×10 mm种植体(巴西圣保罗Conexão®假体系统公司)——EC组(Conexão实验组)。使用放大倍数为408倍的扫描电子显微镜,在每个样本的六个测量部位对种植体-基台界面处的垂直不匹配度进行比较。采用单因素方差分析来检验差异,并使用Tukey检验进行组间两两比较(α = 0.05)。

结果

当Procera基台与其他种植体制造商的产品连接时,发现平均不匹配度存在显著差异。ES组和EC组之间差异不显著,但二者的平均不匹配度均显著高于GNB组(p = 0.001)。GNB组、ES组和EC组的平均不匹配度分别为5.7 µm±0.39、9.53 µm±0.52和10.62 µm±2.16。

结论

Procera氧化锆基台与非其制造商生产的其他种植系统连接时,种植体-基台界面处的垂直不匹配度发生了显著改变。