A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis.

作者信息

Ožiūnas Rimantas, Sakalauskienė Jurgina, Jegelevičius Darius, Janužis Gintaras

机构信息

Department of Prosthodontics, Medical Academy, Lithuanian University of Health Sciences, Kaunas, Lithuania.

Faculty of Electrical and Electronics Engineering, Biomedical Engineering Institute, Kaunas University of Technology, Kaunas, Lithuania.

出版信息

J Adv Prosthodont. 2022 Apr;14(2):70-77. doi: 10.4047/jap.2022.14.2.70. Epub 2022 Apr 27.

Abstract

PURPOSE

This study evaluated screw loosening and 3D crown displacement after cyclic loading of implant-supported incisor crowns cemented with original titanium bases or with three compatible, nonoriginal components.

MATERIALS AND METHODS

A total of 32 dental implants were divided into four groups (n = 8 each): Group 1 used original titanium bases, while Groups 2-4 used compatible components. The reverse torque value (RTV) was evaluated prior to and after cyclic loading (1,200,000 cycles). Samples (prior to and after cyclic loading) were scanned with a microcomputed tomography (micro-CT). Preload and postload files were superimposed by 3D inspection software, and 3D crown displacement analysis was performed using root-mean-square (RMS) values. All datasets were analyzed using one-way ANOVA and Tukey's post hoc analysis.

RESULTS

Significant variations were observed in the postload RTV, depending on the titanium base brand ( < .001). The mean postload RTVs were significantly higher in Groups 1 and 2 than in the other study groups. While evaluating 3D crown displacement, the lowest mean RMS value was shown in the original Group 1, with the highest RMS value occurring in Group 4.

CONCLUSION

Within the limitations of this study and under the implemented conditions, it was concluded that the manufacturer brand of the titanium base significantly influenced screw loosening following the fatigue test and influenced 3D crown displacement after cyclic loading.

摘要

目的

本研究评估了用原装钛基或三种兼容的非原装部件粘结的种植体支持的切牙冠在循环加载后的螺钉松动和三维冠位移情况。

材料与方法

总共32颗牙种植体分为四组(每组n = 8):第1组使用原装钛基,而第2 - 4组使用兼容部件。在循环加载(1,200,000次循环)之前和之后评估反向扭矩值(RTV)。用微型计算机断层扫描(micro-CT)对样本(循环加载之前和之后)进行扫描。通过三维检测软件将加载前和加载后的文件进行叠加,并使用均方根(RMS)值进行三维冠位移分析。所有数据集均使用单因素方差分析和Tukey事后分析进行分析。

结果

根据钛基品牌的不同,加载后的RTV观察到显著差异(P <.001)。第1组和第2组的平均加载后RTV显著高于其他研究组。在评估三维冠位移时,原装的第1组显示出最低的平均RMS值,而第4组的RMS值最高。

结论

在本研究的局限性和实施条件下,得出的结论是,钛基的制造商品牌在疲劳试验后对螺钉松动有显著影响,并在循环加载后对三维冠位移有影响。