A simulator for oscillometric blood-pressure signals to test automated noninvasive sphygmomanometers.

作者信息

Riedel W, Mieke S, Seemann R, Ittermann B

机构信息

Physikalisch-Technische Bundesanstalt, Berlin, Germany.

出版信息

Rev Sci Instrum. 2011 Feb;82(2):024303. doi: 10.1063/1.3549803.

Abstract

A device was developed allowing to generate simulated human blood pressure signals for the purpose of testing the performance of automated noninvasive sphygmomanometers. The apparatus reproducibly generates blood-pressure oscillations synthesized from prerecorded measurements on human subjects. These real-life data allow for a much better evaluation of the accuracy of blood-pressure measurements than the existing simulators using artificial and thus less realistic waveforms. To assess the performance of a given sphygmomanometer under both stable and varying conditions, generated signals can be repeated in their original shape or distorted by well-defined artifacts. In comparison to clinical tests, the procedural influences on the performance testing of sphygmomanometers are largely reduced when the simulator is used.

摘要

开发了一种设备,用于生成模拟人体血压信号,以测试自动无创血压计的性能。该装置可重复生成由对人体受试者预先记录的测量数据合成的血压振荡。与现有的使用人工且不太逼真波形的模拟器相比,这些真实数据能更好地评估血压测量的准确性。为了评估给定血压计在稳定和变化条件下的性能,生成的信号可以按其原始形状重复,或者被明确界定的伪像扭曲。与临床试验相比,使用模拟器时,对血压计性能测试的程序影响在很大程度上得以降低。