The purpose of this study was to evaluate the effects of longitudinal drift in scanner hardware, inter-scanner variability (bias), and scanner upgrade on longitudinal changes in global and regional diffusion properties using longitudinal data obtained on two scanners of the exact same model at one institution. A total of 224 normal subjects were scanned twice, at an interval of about 1 year, using two 3.0-T scanners of the exact same model. Both scanners were simultaneously upgraded during the study period. The subjects were divided into four groups according to the combination of scanners used. With use of tract-based spatial statistics, we evaluated the effects of scanner drift and inter-scanner variability (bias) on global and regional fractional anisotropy (FA), axial diffusivity (AD), and radial diffusivity (RD) changes of the white matter. Even with scanners of the exact same model, inter-scanner variability (bias) significantly affected longitudinal results. FA, AD, and RD of the white matter were relatively stable within the same scanner. We also investigated the effect of scanner upgrade on longitudinal FA, AD, and RD changes. The scanner upgrade included only software upgrade, not hardware upgrade; however, there was a significant effect of scanner upgrade on longitudinal results. These results indicate that inter-scanner variability and scanner upgrade can significantly affect the results of longitudinal diffusion tensor imaging studies.
本研究旨在评估扫描仪硬件的纵向漂移、扫描仪间的变异性(偏差)以及扫描仪升级对使用同一机构两台完全相同型号扫描仪获得的纵向数据的全局和区域扩散性质的纵向变化的影响。共有 224 名正常受试者在大约 1 年的时间内使用两台完全相同型号的 3.0-T 扫描仪进行了两次扫描。在研究期间,两台扫描仪都同时进行了升级。根据使用的扫描仪组合,将受试者分为四组。使用基于束的空间统计学,我们评估了扫描仪漂移和扫描仪间变异性(偏差)对脑白质的全局和区域各向异性分数(FA)、轴向扩散系数(AD)和径向扩散系数(RD)变化的影响。即使使用完全相同型号的扫描仪,扫描仪间变异性(偏差)也会显著影响纵向结果。同一扫描仪内脑白质的 FA、AD 和 RD 相对稳定。我们还研究了扫描仪升级对纵向 FA、AD 和 RD 变化的影响。扫描仪升级仅包括软件升级,不包括硬件升级;然而,扫描仪升级对纵向结果有显著影响。这些结果表明,扫描仪间变异性和扫描仪升级会显著影响纵向扩散张量成像研究的结果。