正畸微种植体稳定性与导孔种植体直径比。

Orthodontic mini-implant stability and the ratio of pilot hole implant diameter.

机构信息

Department of Orthodontics, Nihon University School of Dentistry, 1-8-13 Kandasurugadai, Chiyodaku, Tokyo, Japan.

出版信息

Eur J Orthod. 2012 Feb;34(1):52-6. doi: 10.1093/ejo/cjq157. Epub 2011 Mar 10.

Abstract

One notable complication of mini-implants that are used to provide anchorage in orthodontic treatment is loosening. The aim of this study was to evaluate the relationship between mini-implant mobility during the healing phase and the prognosis for implant stability. Twenty male Wistar rats (aged 20 weeks) were used. Drills with diameters of 0.8, 0.9, 1.0, and 1.1 mm were used to make pilot holes in the rat tibiae. The inserted mini-implants (diameter 1.4 mm; spearhead 1.2 mm; halfway between maximum and minimum 1.3 mm; length 4.0 mm) were subjected to an experimental traction of force for 3 weeks. Bone-to-implant contact (BIC) was observed histologically. Another 20 male rats (aged 20 weeks) underwent an identical procedure, and the stability of the mini-implants was measured using the Periotest before and after traction. The data were statistically analysed using Scheffé's test. The BIC ratios of the 0.9 and 1.0 mm groups were significantly greater than those of the other groups. The Periotest values measured 3 weeks after implant insertion were significantly lower (P < 0.05) than those measured at insertion, except in the 1.1 mm group. To obtain mini-implant stability, the hole diameter should be between 69 and 77 per cent of the diameter of the mini-implant. A significant decrease in the mobility of the mini-implants 3 weeks post-insertion implies a good prognosis for the subsequent mini-implant stability.

摘要

微型种植体在正畸治疗中提供支抗时,一个显著的并发症是松动。本研究旨在评估愈合期微型种植体动度与种植体稳定性预后之间的关系。使用 20 只雄性 Wistar 大鼠(20 周龄)。在大鼠胫骨上用直径为 0.8、0.9、1.0 和 1.1mm 的钻头制作导孔。插入直径为 1.4mm(箭头 1.2mm;最大和最小直径之间 1.3mm;长度 4.0mm)的微型种植体,并施加 3 周的实验牵引力。通过组织学观察骨-种植体接触(BIC)。另外 20 只雄性大鼠(20 周龄)进行了相同的程序,在牵引前后使用 Periotest 测量微型种植体的稳定性。使用 Scheffé 检验对数据进行统计学分析。0.9 和 1.0mm 组的 BIC 比值明显大于其他组。植入后 3 周测量的 Periotest 值明显低于植入时的测量值(P<0.05),但 1.1mm 组除外。为了获得微型种植体的稳定性,孔径应在微型种植体直径的 69%至 77%之间。植入后 3 周微型种植体动度显著降低,预示着随后的微型种植体稳定性良好。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验