Department of Surgery, Plymouth Hospitals NHS Trust, Plymouth, UK.
Br J Surg. 2011 Jun;98(6):797-800. doi: 10.1002/bjs.7433. Epub 2011 Mar 25.
Several instruments are available for open and laparoscopic dissection, including electrothermal bipolar vessel sealers and ultrasonically coagulating shears. The vessel sealing ability of three devices in colorectal specimens was compared in an experimental study.
Surgical specimens from patients scheduled for elective open or laparoscopic colorectal resection were allocated to one of the three devices. After removal of the surgical specimen, up to eight mesenteric vessels were dissected ex vivo and sealed using the allocated instrument. The vessel seal was tested for the maximum pressure at which it leaked and then assessed by a pathologist for depth of thermal tissue damage.
A total of 93 vessels from 18 patients were assessed ex vivo (LOTUS™ n = 33; Harmonic Ace® n = 30; LigaSure™ n = 30), a median of 6 (range 1-8) vessels per surgical specimen with a mean(s.d.) diameter of 1·06(0·70) mm and wall thickness of 0·29(0·19) mm. Mean(s.d.) bursting pressures were 1170(440), 1470(670) and 1510(740) mmHg with LOTUS(™) , Harmonic Ace® and LigaSure™ respectively. ANCOVA showed no difference in bursting pressure between the instruments (P = 0·058). The depth of thermal damage was significantly greater with LigaSure™ (3·37(1·44) mm) than with LOTUS(™) (2·18(0·99) mm; P < 0·001) or Harmonic Ace® (1·95(0·92) mm; P < 0·001).
All three instruments were equally good at sealing blood vessels, with bursting pressures well above physiological blood pressure levels.
NCT01121614 (http://www.clinicaltrials.gov).
有几种器械可用于开放式和腹腔镜下的解剖,包括电热双极血管密封器和超声凝血剪。在一项实验研究中,比较了三种器械在结直肠标本中的血管密封能力。
将择期行开放式或腹腔镜结直肠切除术的患者的手术标本分配至三种器械中的一种。标本切除后,离体解剖 8 条肠系膜血管,并使用分配的器械进行血管密封。测量血管密封泄漏的最大压力,并由病理学家评估热组织损伤的深度。
共对 18 例患者的 93 条血管进行了离体评估(LOTUS ™ n = 33;Harmonic Ace ® n = 30;LigaSure ™ n = 30),每例手术标本有 6 条(范围 1-8 条)血管,平均(标准差)直径为 1.06(0.70)mm,壁厚 0.29(0.19)mm。LOTUS ™ 、Harmonic Ace ® 和 LigaSure ™ 的平均(标准差)破裂压力分别为 1170(440)、1470(670)和 1510(740)mmHg。协方差分析显示,三种器械之间的破裂压力无差异(P = 0.058)。LigaSure ™ 的热损伤深度明显大于 LOTUS ™ (3.37(1.44)mm;P < 0.001)和 Harmonic Ace ® (1.95(0.92)mm;P < 0.001)。
三种器械在密封血管方面同样有效,破裂压力远高于生理血压水平。
NCT01121614(http://www.clinicaltrials.gov)。