Blood--artificial surface interactions during cardiopulmonary bypass. A comparative study of four oxygenators.

作者信息

Benedetti M, De Caterina R, Bionda A, Gardinali M, Cicardi M, Maffei S, Gazzetti P, Pistolesi P, Vernazza F, Michelassi C

机构信息

Department of Cardiac Surgery, University of Pisa, Italy.

出版信息

Int J Artif Organs. 1990 Aug;13(8):488-97.

PMID:2146229
Abstract

Evaluation of the biocompatibility of four different types of oxygenator (bubble, membrane, hollow fibre and 'hybrid') was performed on 26 patients undergoing cardiopulmonary bypass during elective coronary surgery. More platelet derangement and an increased degree of hemolysis, revealed by higher plasmatic concentration of beta-thromboglobulin, platelet factor 4 and plasmatic free hemoglobin (p less than 0.05), was seen when using the bubble oxygenator. Damage to blood cells was minimal with the membrane oxygenator while the 'hybrid' and the hollow fibre oxygenators proved to rank at an intermediate level. Complement activation at the beginning of the cardiopulmonary bypass occurred via the alternative pathway as demonstrated by C3ades arg increase (up to nine times) without a concomitant elevation of C4ades arg. Cardiopulmonary bypass complement activation was quantitatively similar with all the oxygenators. A further activation via the classical pathway occurred in all the patients after protamine injection. Consistent differences as far as clinical and biological effects exist among the various commercially available cardiopulmonary bypass apparatus; our study provides guidelines for the evaluation and selection of devices which might reduce postoperative sequelae.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验