在[18F]FDG生产过程中,对Tracerlab MX系统上一次性盒中留存的长寿命放射性杂质的测量。

Measurement of long lived radioactive impurities retained in the disposable cassettes on the Tracerlab MX system during the production of [18F]FDG.

作者信息

Ferguson D, Orr P, Gillanders J, Corrigan G, Marshall C

机构信息

Regional Medical Physics Service, Belfast Health and Social Care Trust, Belfast, Northern Ireland.

出版信息

Appl Radiat Isot. 2011 Oct;69(10):1479-85. doi: 10.1016/j.apradiso.2011.05.028. Epub 2011 May 30.

Abstract

Using a High- Purity Germanium gamma-ray spectrometer, a number of radioisotopes have been identified within Tracerlab MX radiochemistry system cassettes used to synthesise [18F]FDG. Twenty radiochemistry cassettes were measured and the average total activity of each radioisotope was determined. Using these values and decay correction, the minimum time the cassettes should be left in a decay store before the specific activity falls below 0.4B q/g, the limit for disposal alongside Clinical Waste was found to be 24 months.

摘要

使用高纯锗γ射线光谱仪,在用于合成[18F]FDG的Tracerlab MX放射化学系统盒中鉴定出了多种放射性同位素。对20个放射化学盒进行了测量,并确定了每种放射性同位素的平均总活度。利用这些值和衰变校正,发现放射性比活度降至低于0.4Bq/g(与临床废物一起处置的限值)之前,这些盒应留在衰变储存库中的最短时间为24个月。