Ferguson D, Orr P, Gillanders J, Corrigan G, Marshall C
Regional Medical Physics Service, Belfast Health and Social Care Trust, Belfast, Northern Ireland.
Appl Radiat Isot. 2011 Oct;69(10):1479-85. doi: 10.1016/j.apradiso.2011.05.028. Epub 2011 May 30.
Using a High- Purity Germanium gamma-ray spectrometer, a number of radioisotopes have been identified within Tracerlab MX radiochemistry system cassettes used to synthesise [18F]FDG. Twenty radiochemistry cassettes were measured and the average total activity of each radioisotope was determined. Using these values and decay correction, the minimum time the cassettes should be left in a decay store before the specific activity falls below 0.4B q/g, the limit for disposal alongside Clinical Waste was found to be 24 months.