Contactless conductivity detection for capillary electrophoresis.

作者信息

Zemann A J, Schnell E, Volgger D, Bonn G K

机构信息

Institute of Analytical Chemistry and Radiochemistry, Leopold-Franzens-University Innsbruck, Innrain 52a, A-6020 Innsbruck, Austria.

出版信息

Anal Chem. 1998 Feb 1;70(3):563-7. doi: 10.1021/ac9707592.

Abstract

A contactless capacitively coupled conductivity detector for capillary electrophoresis is introduced. The detector consists of two electrodes which are placed cylindrically around the outer polyimide coating of the fused-silica capillary with a detection gap of 2 mm. The electrodes form a cylindrical capacitor, and the electric conductivity of the solution in the gap between the electrodes is measured. A high audio or low ultrasonic frequency for coupling of the ac voltage is used in order to minimize the influence of reactance of the liquid. For an improved version of the detector, two syringe cannulas are used as the electrodes and the capillary is simply assembled into the tubing. This allows an easy placement of the detector on various positions along the capillary. The limit of detection of inorganic cations and anions is 200 ppb, as determined for sodium and chloride, respectively.

摘要

介绍了一种用于毛细管电泳的非接触式电容耦合电导检测器。该检测器由两个电极组成,它们围绕熔融石英毛细管的外部聚酰亚胺涂层呈圆柱形放置,检测间隙为2毫米。电极形成一个圆柱形电容器,并测量电极之间间隙中溶液的电导率。为了使液体电抗的影响最小化,使用高音频或低超声频率来耦合交流电压。对于该检测器的改进版本,使用两个注射器套管作为电极,毛细管简单地组装到 tubing 中。这使得检测器能够轻松地放置在沿毛细管的各个位置。无机阳离子和阴离子的检测限分别为200 ppb,这是针对钠和氯测定的。