Suppr超能文献

英国高血压学会关于评估自动和半自动血压测量设备(特别涉及动态血压监测系统)的方案。

The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems.

作者信息

O'Brien E, Petrie J, Littler W, de Swiet M, Padfield P L, O'Malley K, Jamieson M, Altman D, Bland M, Atkins N

机构信息

Blood Pressure Unit, Beaumont Hospital, Dublin, Ireland.

出版信息

J Hypertens. 1990 Jul;8(7):607-19. doi: 10.1097/00004872-199007000-00004.

Abstract

With the increasing manufacture of expensive systems for the measurement of ambulatory blood pressure there is a need for potential purchasers to be able to satisfy themselves that the systems have been evaluated according to agreed criteria. The British Hypertension Society has, therefore, drawn up a protocol of requirements for the evaluation of these devices. This protocol incorporates many features of the American National Standard for Non-Automated Sphygmomanometers but includes many additional features, such as strict criteria for observer training, interdevice variability testing before and after a month of ambulatory use, and a new system of analysis which permits the test system to be graded. It is recommended that manufacturers of ambulatory blood pressure measuring devices should obtain an unbiased evaluation according to a recognized standard before a device is marketed.

摘要

随着用于测量动态血压的昂贵系统的生产日益增多,潜在购买者需要能够确信这些系统已根据商定的标准进行了评估。因此,英国高血压学会制定了这些设备评估的要求方案。该方案纳入了美国非自动血压计国家标准的许多特点,但也包括许多额外的特点,如严格的观察者培训标准、动态使用一个月前后的设备间变异性测试,以及一种允许对测试系统进行分级的新分析系统。建议动态血压测量设备的制造商在设备上市前应根据公认标准获得无偏见的评估。