评估用于放射治疗计划的平板床采集的盆腔磁共振图像的质量。

Assessing the image quality of pelvic MR images acquired with a flat couch for radiotherapy treatment planning.

机构信息

Radiotherapy Physics Department, Northern Ireland Cancer Centre, Belfast, Northern Ireland, UK.

出版信息

Br J Radiol. 2011 Aug;84(1004):750-5. doi: 10.1259/bjr/27295679.

Abstract

OBJECTIVES

To improve the integration of MRI with radiotherapy treatment planning, our department fabricated a flat couch top for our MR scanner. Setting up using this couch top meant that the patients were physically higher up in the scanner and, posteriorly, a gap was introduced between the patient and radiofrequency coil.

METHODS

Phantom measurements were performed to assess the quantitative impact on image quality. A phantom was set up with and without the flat couch insert in place, and measurements of image uniformity and signal to noise were made. To assess clinical impact, six patients with pelvic cancer were recruited and scanned on both couch types. The image quality of pairs of scans was assessed by two consultant radiologists.

RESULTS

The use of the flat couch insert led to a drop in image signal to noise of approximately 14%. Uniformity in the anteroposterior direction was affected the most, with little change in right-to-left and feet-to-head directions. All six patients were successfully scanned on the flat couch, although one patient had to be positioned with their arms by their sides. The image quality scores showed no statistically significant change between scans with and without the flat couch in place.

CONCLUSION

Although the quantitative performance of the coil is affected by the integration of a flat couch top, there is no discernible deterioration of diagnostic image quality, as assessed by two consultant radiologists. Although the flat couch insert moved patients higher in the bore of the scanner, all patients in the study were successfully scanned.

摘要

目的

为了提高 MRI 与放射治疗计划的整合度,我们部门为我们的磁共振扫描仪制作了一个平板床面。使用这个床面意味着患者在扫描仪中的位置更高,并且在患者和射频线圈之间会产生一个间隙。

方法

进行了体模测量,以评估图像质量的定量影响。在有和没有平板床面插入的情况下设置了一个体模,并测量了图像均匀性和信噪比。为了评估临床影响,招募了 6 名患有骨盆癌的患者,并在两种床面类型上进行了扫描。两位顾问放射科医生对成对扫描的图像质量进行了评估。

结果

使用平板床面插入导致图像信噪比下降约 14%。前后方向的均匀性受影响最大,左右和头尾方向的变化较小。所有 6 名患者都成功地在平板床上进行了扫描,尽管有 1 名患者必须将手臂放在身体两侧进行定位。图像质量评分显示,在有和没有平板床面的情况下,扫描之间没有统计学上的显著变化。

结论

尽管平板床面的集成会影响线圈的定量性能,但经过两位顾问放射科医生评估,诊断图像质量没有明显恶化。尽管平板床面插入将患者移到了扫描仪的孔内更高的位置,但研究中的所有患者都成功进行了扫描。

相似文献

10
Reducing patient posture variability using the predicted couch position.使用预测的治疗床位置减少患者体位变异性。
Med Dosim. 2015 Autumn;40(3):218-21. doi: 10.1016/j.meddos.2014.12.002. Epub 2015 Jan 22.

引用本文的文献

2
The use of PET/MRI in radiotherapy.正电子发射断层显像/磁共振成像在放射治疗中的应用。
Insights Imaging. 2024 Feb 27;15(1):63. doi: 10.1186/s13244-024-01627-6.

本文引用的文献

3
Dedicated magnetic resonance imaging in the radiotherapy clinic.放疗临床专用磁共振成像
Int J Radiat Oncol Biol Phys. 2009 Jun 1;74(2):644-51. doi: 10.1016/j.ijrobp.2009.01.065.
4
Anatomical imaging for radiotherapy.放射治疗的解剖成像
Phys Med Biol. 2008 Jun 21;53(12):R151-91. doi: 10.1088/0031-9155/53/12/R01. Epub 2008 May 21.
10
MR image-guided portal verification for brain treatment field.用于脑部治疗区域的磁共振成像引导的射野验证
Int J Radiat Oncol Biol Phys. 1998 Feb 1;40(3):703-11. doi: 10.1016/s0360-3016(97)00763-3.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验