FOM Institute for Atomic and Molecular Physics (AMOLF), Science Park 104, 1098 XG Amsterdam, The Netherlands.
Anal Chem. 2011 Oct 15;83(20):7888-94. doi: 10.1021/ac2017629. Epub 2011 Sep 22.
Highly parallel, active pixel detectors enable novel detection capabilities for large biomolecules in time-of-flight (TOF) based mass spectrometry imaging (MSI). In this work, a 512 × 512 pixel, bare Timepix assembly combined with chevron microchannel plates (MCP) captures time-resolved images of several m/z species in a single measurement. Mass-resolved ion images from Timepix measurements of peptide and protein standards demonstrate the capability to return both mass-spectral and localization information of biologically relevant analytes from matrix-assisted laser desorption ionization (MALDI) on a commercial ion microscope. The use of a MCP-Timepix assembly delivers an increased dynamic range of several orders of magnitude. The Timepix returns defined mass spectra already at subsaturation MCP gains, which prolongs the MCP lifetime and allows the gain to be optimized for image quality. The Timepix peak resolution is only limited by the resolution of the in-pixel measurement clock. Oligomers of the protein ubiquitin were measured up to 78 kDa.
高度并行的有源像素探测器使基于飞行时间 (TOF) 的质谱成像 (MSI) 能够对大型生物分子进行新的检测。在这项工作中,一个 512×512 像素的裸 Timepix 组件与 V 形微通道板 (MCP) 结合,可在单次测量中捕获多个 m/z 物种的时间分辨图像。Timepix 对肽和蛋白质标准品的测量得到的质谱分辨离子图像证明了从基质辅助激光解吸电离 (MALDI) 中返回商业离子显微镜中生物相关分析物的质谱和定位信息的能力。MCP-Timepix 组件的使用提供了几个数量级的动态范围增加。Timepix 在亚饱和 MCP 增益下已经返回了定义明确的质谱,这延长了 MCP 的寿命,并允许根据图像质量优化增益。Timepix 的峰值分辨率仅受像素内测量时钟分辨率的限制。我们测量了高达 78 kDa 的蛋白质泛素低聚物。