根据2002年欧洲高血压学会方案和英国高血压学会方案评估Microlife WatchBP Office ABI监测仪的准确性。

Accuracy of Microlife WatchBP Office ABI monitor assessed according to the 2002 European Society of Hypertension protocol and the British Hypertension Society protocol.

作者信息

Saladini Francesca, Benetti Elisabetta, Masiero Serena, Palatini Paolo

机构信息

Department of Clinical and Experimental Medicine, University of Padua, Padua, Italy.

出版信息

Blood Press Monit. 2011 Oct;16(5):258-61. doi: 10.1097/MBP.0b013e32834af72e.

Abstract

OBJECTIVE

To determine the accuracy of the WatchBP Office ABI monitor for blood pressure measurement developed by the Microlife Company.

METHODS

The device accuracy was tested in 85 subjects with a mean age of 54 ± 19 years. Their systolic and diastolic blood pressure (SBP/DBP) at entry was 141 ± 30/86 ± 19 mmHg, and upper arm circumference was 28 ± 5 cm. Initially, the data from 33 participants were examined according to the 2002 version of the European Society of Hypertension (ESH) protocol. An additional 52 subjects were then enrolled to fulfill the requirements of the British Hypertension Society (BHS) protocol. In all participants, sequential same arm measurements were performed by two trained observers.

RESULTS

The device passed all three phases of the ESH protocol for SBP and DBP. For the BHS protocol the device was graded A for both SBP and DBP. The A/A grade was achieved in the low blood pressure category (< 130/80 mmHg), the B/A grade in the medium category (130-160/80-100 mmHg) and the A/A grade in the high category (> 160/100 mmHg). Mean blood pressure difference between device and observers in the first 33 subjects was -0.9 ± 5.5 mmHg for SBP and -2.2 ± 4.5 mmHg for DBP and in the 85 participants it was -1.2 ± 6.5 mmHg and -2.3 ± 5.1, respectively.

CONCLUSION

These data show that the Microlife WatchBP Office ABI monitor satisfied the recommended ESH accuracy levels and achieved A/A grade of the BHS protocol across a wide range of BP.

摘要

目的

确定Microlife公司研发的WatchBP Office ABI血压监测仪测量血压的准确性。

方法

对85名平均年龄为54±19岁的受试者进行了设备准确性测试。他们入组时的收缩压和舒张压(SBP/DBP)为141±30/86±19 mmHg,上臂围为28±5 cm。最初,根据2002版欧洲高血压学会(ESH)方案对33名参与者的数据进行了检查。随后又招募了52名受试者以满足英国高血压学会(BHS)方案的要求。在所有参与者中,由两名经过培训的观察者对同一侧手臂进行连续测量。

结果

该设备在ESH方案的SBP和DBP三个阶段均通过测试。对于BHS方案,该设备的SBP和DBP均被评为A级。在低血压类别(<130/80 mmHg)中达到A/A级,在中等血压类别(130 - 160/80 - 100 mmHg)中达到B/A级,在高血压类别(>160/100 mmHg)中达到A/A级。在前33名受试者中,设备与观察者之间的平均血压差值SBP为-0.9±5.5 mmHg,DBP为-2.2±4.5 mmHg;在85名参与者中,该差值SBP为-1.2±6.5 mmHg,DBP为-2.3±5.1 mmHg。

结论

这些数据表明,Microlife WatchBP Office ABI监测仪满足ESH推荐的准确性水平,并在广泛的血压范围内达到了BHS方案的A/A级。