Suppr超能文献

应用压平眼压计和 Tono-pen 测量犬眼的角膜模量和眼压。

Corneal modulus and IOP measurements in canine eyes using Goldmann applanation tonometry and Tono-pen.

机构信息

Department of Biomedical Engineering, Ohio State University, Columbus, Ohio, USA.

出版信息

Invest Ophthalmol Vis Sci. 2011 Oct 4;52(11):7866-71. doi: 10.1167/iovs.11-7407.

Abstract

PURPOSE

To experimentally examine the effect of corneal modulus on Goldmann applanation tonometry (GAT) and Tono-pen (Tono-pen XL, Reichert, Inc., Depew, NY) measurements of intraocular pressure (IOP) in a canine eye model.

METHODS

Twenty-one canine globes were recovered from healthy animals. IOP was controlled at 10, 15, 20, 30, and 40 mm Hg and measured by GAT and Tono-pen following standard protocols. The corneas were dissected and uniaxial tensile tests were performed on corneal strips. The correlation between GAT and Tono-pen errors and corneal secant modulus was evaluated using Pearson correlation coefficients. The influence of corneal thickness and the true pressure was also examined.

RESULTS

At a true IOP of 10, 15, 20, 30, and 40 mm Hg, the GAT readings were 1.1 ± 1.0, 5.1 ± 1.5, 9.5 ± 2.0, 17.3 ± 1.6, and 25.3 ± 1.8 mm Hg, respectively. The corresponding Tono-pen readings were 7.8 ± 1.7, 12.4 ± 1.7, 16.1 ± 1.9, 22.5 ± 2.1, and 28.1 ± 2.2 mm Hg, respectively. The mean secant modulus at 1% strain of the canine corneal strips was 1.54 ± 0.43 megapascal (MPa). Corneal secant modulus was significantly correlated with GAT errors when the true IOP was 30 mm Hg (R = 0.49; P < 0.05). No significant correlation was observed between tonometric errors and corneal thickness. Both GAT and Tono-pen errors increased significantly at higher pressures (P < 0.001).

CONCLUSIONS

Both GAT and Tono-pen underestimated IOP in canine eyes. There was preliminary experimental evidence for a correlation between corneal modulus and GAT in the canine eyes and a higher corneal modulus was associated with higher GAT readings at a certain pressure level. The tonometric errors appeared to be pressure-dependent.

摘要

目的

在犬眼模型中,通过实验研究角膜模量对 Goldmann 压平眼压计(GAT)和 Tono-pen(Tono-pen XL,Reichert,Inc.,Depew,NY)眼压测量的影响。

方法

从健康动物中回收 21 个犬眼球。眼压控制在 10、15、20、30 和 40 mmHg,按照标准方案通过 GAT 和 Tono-pen 进行测量。对角膜条进行解剖并进行单轴拉伸试验。使用 Pearson 相关系数评估 GAT 和 Tono-pen 误差与角膜割线模量之间的相关性。还检查了角膜厚度和真实压力的影响。

结果

在真实眼压为 10、15、20、30 和 40 mmHg 时,GAT 读数分别为 1.1 ± 1.0、5.1 ± 1.5、9.5 ± 2.0、17.3 ± 1.6 和 25.3 ± 1.8 mmHg。相应的 Tono-pen 读数分别为 7.8 ± 1.7、12.4 ± 1.7、16.1 ± 1.9、22.5 ± 2.1 和 28.1 ± 2.2 mmHg。犬角膜条在 1%应变时的平均割线模量为 1.54 ± 0.43 兆帕斯卡(MPa)。当真实眼压为 30 mmHg 时,角膜割线模量与 GAT 误差显著相关(R = 0.49;P < 0.05)。眼压计误差与角膜厚度之间无显著相关性。随着眼压的升高,GAT 和 Tono-pen 的误差均显著增加(P < 0.001)。

结论

GAT 和 Tono-pen 均低估了犬眼的眼压。在犬眼中,有初步的实验证据表明角膜模量与 GAT 之间存在相关性,并且在一定压力水平下,角膜模量较高与 GAT 读数较高相关。眼压计误差似乎与压力有关。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验