Biosens Bioelectron. 2012 Jan 15;31(1):26-31. doi: 10.1016/j.bios.2011.09.048. Epub 2011 Oct 20.
In this study, a novel method was proposed for the selective determination of tryptophan (TRP) in blood serum in the presence of tyrosine. This method is based on the electrochemical reduction of 2-amino-3-(5-oxo-3,5-dihydro-2H-indol-3-yl)-propionic acid (5-O-3,5DH-TRP) formed by the oxidation of TRP on the electrochemically treated pencil graphite (ETPG) electrode surface at a suitable potential value. The parameters affecting the TRP determination were deeply investigated. The optimal pH value was determined as 3. The highest reduction current intensity was obtained at the accumulation potential and time values of +0.95 V and 120 s, respectively. The reduction peak current values of 5-O-3,5DH-TRP versus TRP concentration at the ETPG electrode showed linearity in the range from 0.5 μM to 50.0 μM (R(2)=0.9962) with a detection limit of 0.05 μM (S/N=3). The reduction peak intensity of 5-O-3,5DH-TRP on the ETPG electrode showed no significant change in the presence of different interfering substances. The analytical application of the proposed novel method was successfully tested by using human blood serum samples.
在这项研究中,提出了一种在酪氨酸存在下选择性测定血清中天冬氨酸(TRP)的新方法。该方法基于在电化学处理的铅笔石墨(ETPG)电极表面上合适的电位值下氧化 TRP 形成的 2-氨基-3-(5-氧代-3,5-二氢-2H-吲哚-3-基)-丙酸(5-O-3,5DH-TRP)的电化学还原。深入研究了影响 TRP 测定的参数。确定最佳 pH 值为 3。在+0.95 V 和 120 s 的积累电位和时间值下,获得了最高的还原电流强度。在 ETPG 电极上,5-O-3,5DH-TRP 的还原峰电流值与 TRP 浓度在 0.5 μM 至 50.0 μM 范围内呈线性关系(R²=0.9962),检测限为 0.05 μM(S/N=3)。在存在不同干扰物质的情况下,ETPG 电极上 5-O-3,5DH-TRP 的还原峰强度没有明显变化。该新方法的分析应用已成功通过使用人血清样品进行了测试。