用于循环性休克和呼吸衰竭患者的三根混合静脉血氧饱和度导管。

Three mixed venous saturation catheters in patients with circulatory shock and respiratory failure.

作者信息

Rouby J J, Poète P, Bodin L, Bourgeois J L, Arthaud M, Viars P

机构信息

Départment d'Anesthésie de la Pitié, Université Paris VI, France.

出版信息

Chest. 1990 Oct;98(4):954-8. doi: 10.1378/chest.98.4.954.

Abstract

Thirty-one critically ill patients with acute respiratory failure and circulatory shock were divided into three groups. Group 1 included 11 patients with an inserted mixed venous saturation (SvO2) catheter using three-reference wavelengths and characterized by one transmitting and one detecting fiberoptic filament (Oximetrix opticath catheter); group 2 included eight patients with an inserted SvO2 catheter using two-reference wavelengths and characterized by one transmitting and one detecting fiberoptic filament; (Edwards sat-one catheter); group 3 included 12 patients with an inserted SvO2 catheter using two-reference wavelengths and characterized by one transmitting and two detecting fiberoptic filaments. Once calibration procedures were performed, SvO2 measured by the catheter and by an hemoximeter OSM 3 (reference value) were compared following each therapeutic intervention. Over a period of 1.5 to 6 hours during which the hematocrit value remained unchanged, 119 measurements were obtained in group 1, 91 in group 2 and 181 in group 3. The dispersion of SvO2 values was much more pronounced with the two-reference wavelength systems using either one or two detecting fiber optic filaments, and the correlation coefficient was significantly higher with the three-reference wavelength system (r = 0.970 for the Oximetrix catheter vs r = 0.855 for the Edwards catheter and r = 0.826 for the Spectramed catheter, p less than 0.001). After 24 hours, the spontaneous drifts in the two-reference wavelength systems, using either one or two detecting fiber optic filaments (expressed as the SvO2 value measured by the catheter minus the reference SvO2 value) were significantly higher than the spontaneous drift in the three-reference wavelength system (9.3 +/- 7 percent for the Edwards catheter and +/- 6 +/- 4.1 percent for the Spectramed catheter vs 3.3 +/- 3.1 percent for the Oximetrix catheter, p less than 0.05). This study shows that a three-wavelength system is more accurate than a two-wavelength system for measuring acute changes in SvO2. The addition of a second detecting fiber optic filament does not seem to improve the accuracy of the system when SvO2 changes occur in conditions of stable hematocrit.

摘要

31例患有急性呼吸衰竭和循环性休克的重症患者被分为三组。第一组包括11例插入了采用三参考波长的混合静脉血氧饱和度(SvO2)导管的患者,其特征为一根传输光纤丝和一根检测光纤丝(Oximetrix opticath导管);第二组包括8例插入了采用两参考波长的SvO2导管的患者,其特征为一根传输光纤丝和一根检测光纤丝(Edwards sat-one导管);第三组包括12例插入了采用两参考波长的SvO2导管的患者,其特征为一根传输光纤丝和两根检测光纤丝。校准程序完成后,在每次治疗干预后,对导管测量的SvO2值与血氧计OSM 3测量的SvO2值(参考值)进行比较。在血细胞比容值保持不变的1.5至6小时期间,第一组获得了119次测量值,第二组获得了91次,第三组获得了181次。使用一根或两根检测光纤丝的两参考波长系统中,SvO2值的离散度更为明显,而三参考波长系统的相关系数显著更高(Oximetrix导管的r = 0.970,Edwards导管的r = 0.855,Spectramed导管的r = 0.826,p < 0.001)。24小时后,使用一根或两根检测光纤丝的两参考波长系统中的自发漂移(表示为导管测量的SvO2值减去参考SvO2值)显著高于三参考波长系统中的自发漂移(Edwards导管为9.3±7%,Spectramed导管为±6±4.1%,而Oximetrix导管为3.3±3.1%,p < 0.05)。本研究表明,在测量SvO2的急性变化时,三波长系统比两波长系统更准确。当血细胞比容稳定的情况下发生SvO2变化时,添加第二根检测光纤丝似乎并不能提高系统的准确性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验