对危重症患者的混合静脉血氧饱和度进行持续监测。

Continuous monitoring of mixed venous oxygen saturation in critically ill patients.

作者信息

Baele P L, McMichan J C, Marsh H M, Sill J C, Southorn P A

出版信息

Anesth Analg. 1982 Jun;61(6):513-7.

PMID:7200741
Abstract

A new pulmonary artery balloon flow-directed catheter combines a fiberoptic photometric system for continuous display of mixed venous blood oxygen saturation (SvO2) with the capacity for hemodynamic measurements including thermodilution cardiac output estimation. This oximetry system was studied to determine its accuracy, reliability, and usefulness in the surgical intensive care unit (ICU). Twenty-two catheters were tested, but only 17 were successfully placed in 16 patients. There were technical problems associated with 10 catheters and on six occasions these necessitated the use of another catheter. The catheter values for SvO2 were closely related (r = 0.9516) to those obtained from a laboratory Co-oximeter. Continuous monitoring of SvO2 is accurate and valuable as a warning system for deterioration in cardiopulmonary function and as an indicator of the effects of various therapeutic maneuvers in critically ill patients.

摘要

一种新型肺动脉气囊漂浮导管,将用于连续显示混合静脉血氧饱和度(SvO2)的光纤光度系统与包括热稀释法心输出量估算在内的血流动力学测量功能相结合。对该血氧测定系统进行了研究,以确定其在外科重症监护病房(ICU)中的准确性、可靠性和实用性。测试了22根导管,但仅17根成功置入16例患者体内。10根导管存在技术问题,有6次因此需要使用另一根导管。导管测得的SvO2值与实验室用的血液气体分析仪测得的值密切相关(r = 0.9516)。连续监测SvO2作为心肺功能恶化的预警系统以及危重症患者各种治疗措施效果的指标,是准确且有价值的。