Department of Oral and Maxillofacial Surgery, Hannover Medical School, Hannover, Germany.
Clin Oral Implants Res. 2012 Oct;23(10):1210-6. doi: 10.1111/j.1600-0501.2011.02291.x. Epub 2011 Nov 14.
The purpose of this study was to evaluate the biomechanical bone tissue response to novel microstructured zirconia implants in comparison to sandblasted and acid-etched (SLA) titanium implants through the analysis of removal torque (RTQ) measurements.
Ti-SLA implants with a sandblasted, large-grit and acid-etched surface were compared with zirconia implants with an acid-etched surface. All implants had the same shape, a diameter of 4.1 mm and a length of 10 mm. A total of 136 implants were placed in the maxillae of 17 miniature pigs. Six animals were sacrificed after both 4 and 8 weeks and five animals were sacrificed after 12 weeks, thus providing a total of 102 implants for RTQ testing (34 implants were reserved for future histological analysis). The RTQ analysis was successfully performed, using a mixed model regression with P-values calculated using the nonparametric Brunner-Langer method, on 100 of the 102 implants, two were excluded from the analysis.
The adjusted mean RTQ values for Ti-SLA implants were 131 Ncm (95% CI: 107-155) at 4 weeks, 128 Ncm (108-148) Ncm at 8 weeks, and 180 Ncm (153-207 Ncm) at 12 weeks of healing, whereas RTQ values for the zirconia implants were 110 Ncm (86-134), 97 Ncm (76-118) and 147 Ncm (121-174) at the corresponding time intervals. A comparison of the implant materials resulted in P-values of P = 0.114 at 4 weeks, P = 0.034 at 8 weeks and P = 0.105 at 12 weeks (significance set at P < 0.05).
Within the limits of the present study, it could be confirmed that the biomechanical bone-tissue response of the investigated zirconia implants is non-inferior to that of the well-documented, roughened titanium surface, at each time point, within the set tolerance. There were no statistically significant differences between the two materials after a healing period of 4 and 12 weeks. The RTQ values of both implant types increased significantly from the 8-week to the 12-week time point.
本研究旨在通过分析取出扭矩(RTQ)测量值,评估新型微结构氧化锆种植体与喷砂酸蚀(SLA)钛种植体的骨组织生物力学反应。
将具有喷砂、大颗粒和酸蚀表面的 Ti-SLA 种植体与具有酸蚀表面的氧化锆种植体进行比较。所有种植体形状相同,直径为 4.1mm,长度为 10mm。总共将 136 个种植体植入 17 只小型猪的上颌骨中。6 只动物分别在 4 周和 8 周后处死,5 只动物在 12 周后处死,因此共有 102 个种植体进行 RTQ 测试(34 个种植体留作未来的组织学分析)。使用混合模型回归分析,对 100 个种植体中的 RTQ 分析进行了成功的分析,其中 2 个种植体被排除在分析之外。采用非参数 Brunner-Langer 方法计算 P 值。
在愈合的 4 周、8 周和 12 周时,Ti-SLA 种植体的调整后平均 RTQ 值分别为 131Ncm(95%CI:107-155)、128Ncm(108-148)和 180Ncm(153-207Ncm),而氧化锆种植体的 RTQ 值分别为 110Ncm(86-134)、97Ncm(76-118)和 147Ncm(121-174)。对种植材料的比较得出的 P 值分别为 4 周时 P=0.114,8 周时 P=0.034,12 周时 P=0.105(显著水平设为 P<0.05)。
在本研究的限制范围内,可以确认在设定的容差内,每个时间点,所研究的氧化锆种植体的骨组织生物力学反应与经过充分验证的粗糙钛表面相当,在愈合 4 周和 12 周后,两种材料之间无统计学差异。两种种植体类型的 RTQ 值均从 8 周时显著增加到 12 周时。