Sir Peter Mansfield Magnetic Resonance Centre, University of Nottingham, Nottingham, NG7 2RD, UK.
Eur J Radiol. 2013 May;82(5):719-27. doi: 10.1016/j.ejrad.2011.09.023. Epub 2011 Dec 3.
T₂*-weighted magnetic resonance imaging at 7 T has recently been shown to allow differentiation between white-matter multiple sclerosis lesions and asymptomatic white-matter lesions, by the presence or absence of a detectable central blood vessel. The aim of the present work is to improve the technique by increasing the sensitivity to veins at both 3 T and 7 T, and to assess the benefit of ultra-high-field imaging. Signal-to-noise ratio (SNR) measurements and simulations are used to compare the sensitivity of magnitude T₂*-weighted and susceptibility-weighted images for the detection of small veins (<1 pixel in diameter), both with and without the use of gadolinium. The simulations are used to predict the optimal scanning parameters in order to increase the sensitivity to these veins at both field strengths, and to reduce the inherent dependence on vessel orientation. The sensitivities of the sequences at both field strengths are compared, theoretically and experimentally, in order to quantify the benefit of imaging at ultra-high-field. Subjects with multiple sclerosis (MS) are scanned at both field strengths, using the optimised sequence parameters, as well as those used in previously published work, and the optimisation is shown to improve the detection of veins within lesions.
7T T₂*-weighted 磁共振成像最近已被证明可通过存在或不存在可检测的中央血管来区分多发性硬化症的脑白质病变和无症状性脑白质病变。本研究旨在通过提高对 3T 和 7T 下静脉的敏感性来改进该技术,并评估超高场成像的优势。使用信噪比(SNR)测量和模拟来比较大小 T₂*-weighted 和磁化率加权图像对小静脉(直径<1 像素)的检测敏感性,包括有无钆剂的情况。模拟用于预测最佳扫描参数,以提高在两种场强下对这些静脉的敏感性,并减少对血管方向的固有依赖性。从理论和实验上比较了两种场强下序列的敏感性,以量化超高场成像的优势。使用优化的序列参数以及之前发表的工作中使用的序列参数,对多发性硬化症(MS)患者进行了两种场强的扫描,结果表明优化可以提高病变内静脉的检测。