To evaluate the repeatability and reproducibility of posterior corneal curvature and posterior corneal elevation best-fit sphere (BFS) obtained with the Visante Omni (Carl Zeiss Meditec) and to compare the results with the Orbscan II (Bausch & Lomb).
Thirty eyes from 30 healthy volunteers were included in this study. All patients were examined 5 times with the Visante Omni and Orbscan II by 2 independent operators. The posterior corneal curvature (3- and 6-mm zone) and posterior corneal elevation BFS (5- and 8-mm zone) were generated for each system. Intraoperator repeatability and interoperator reproducibility and agreement between the systems were evaluated using the intraclass correlation coefficient (ICC) and the Bland-Altman method.
The repeatability of posterior corneal curvature and posterior corneal elevation BFS measured by the Visante Omni was high for all analysis zones (ICC, 0.99 to 1.00). The reproducibility also showed similar results (ICC, 0.99 to 1.00). Agreement between the Visante Omni and Orbscan II was high for posterior corneal curvature (ICC, 0.94 to 0.97) and posterior corneal elevation BFS (ICC, 0.96 to 0.98) with 95% limits of agreement at -0.26 to 0.22 diopters for posterior corneal curvature and 0.11 to 0.69 mm for posterior corneal elevation BFS.
The Visante Omni provides good repeatability and reproducibility of posterior corneal topography. Overall agreement with the Orbscan II system was high.
评估 Visante Omni(卡尔蔡司医学技术)获得的后角膜曲率和后角膜隆起最佳拟合球(BFS)的重复性和可再现性,并与 Orbscan II(博士伦)的结果进行比较。
本研究纳入 30 名健康志愿者的 30 只眼。所有患者均由 2 名独立操作者使用 Visante Omni 和 Orbscan II 检查 5 次。为每个系统生成后角膜曲率(3 和 6mm 区)和后角膜隆起 BFS(5 和 8mm 区)。使用组内相关系数(ICC)和 Bland-Altman 方法评估系统内操作者重复性、操作者间可再现性和系统间一致性。
Visante Omni 测量的后角膜曲率和后角膜隆起 BFS 的重复性在所有分析区均较高(ICC,0.99 至 1.00)。可再现性也显示出类似的结果(ICC,0.99 至 1.00)。Visante Omni 和 Orbscan II 在后角膜曲率(ICC,0.94 至 0.97)和后角膜隆起 BFS(ICC,0.96 至 0.98)方面具有较高的一致性,后角膜曲率的 95%一致性界限为-0.26 至 0.22 屈光度,后角膜隆起 BFS 的一致性界限为 0.11 至 0.69mm。
Visante Omni 在后角膜地形图方面具有良好的重复性和可再现性。与 Orbscan II 系统的总体一致性较高。