快速示波法测定血压的准确性。

The accuracy of rapid oscillometric blood pressure determination.

作者信息

Gorback M S, Quill T J, Graubert D A

机构信息

Department of Anesthesiology, Duke University Medical Center, Durham, North Carolina.

出版信息

Biomed Instrum Technol. 1990 Sep-Oct;24(5):371-4.

PMID:2224342
Abstract

The Dinamap 1846SX (DIN) offers a "stat mode" (SM) for rapid repetitive blood pressure determinations, but its accuracy has not been reported. Thirty patients who required an intra-arterial catheter (IAC) for management of anesthesia underwent induction with concomitant noninvasive blood pressure measurement provided by a DIN set in SM. Computerized automated data acquisition recorded digital outputs from both monitors at the time of new SM readings. Systolic pressure mean absolute error (MAE) was 13.3 +/- 5.7 mmHg (mean +/- SEM) and the mean error (bias) was -8.7 +/- 9.8 mmHg. Mean pressure MAE was 6.8 +/- 3.4 mmHg and bias was -0.5 +/- 6.3 mmHg. Diastolic pressure MAE was 7.0 +/- 3.3 mmHg and bias was -0.3 +/- 6.5 mmHg. There was a tendency for systolic pressure readings to underestimate IAC pressures at higher pressures. These findings echo results reported previously for DIN operating in its normal mode. Rapid arterial pressure determination by SM does not appear to compromise accuracy.

摘要

Dinamap 1846SX(DIN)提供了一种“静态模式”(SM)用于快速重复测量血压,但其准确性尚未见报道。30例因麻醉管理需要置入动脉内导管(IAC)的患者在诱导麻醉时,使用设置为SM模式的DIN进行无创血压测量。计算机自动数据采集记录了每次新的SM读数时两个监测仪的数字输出。收缩压平均绝对误差(MAE)为13.3±5.7 mmHg(平均值±标准误),平均误差(偏差)为-8.7±9.8 mmHg。平均压MAE为6.8±3.4 mmHg,偏差为-0.5±6.3 mmHg。舒张压MAE为7.0±3.3 mmHg,偏差为-0.3±6.5 mmHg。在较高压力下,收缩压读数有低估IAC压力的趋势。这些结果与之前报道的DIN在正常模式下的操作结果一致。SM模式下快速测定动脉压似乎不会影响准确性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验