袖带手臂对示波法血压测量的影响。

The effect of sleeved arms on oscillometric blood pressure measurement.

作者信息

Holleman D R, Westman E C, McCrory D C, Simel D L

机构信息

Center for Health Services Research in Primary Care, Durham Veterans Affairs Medical Center, North Carolina.

出版信息

J Gen Intern Med. 1993 Jun;8(6):325-6. doi: 10.1007/BF02600148.

Abstract

The authors measured the blood pressures of 36 subjects who had bare and sleeved arms to determine the effect of wearing sleeves on automatic oscillometric blood pressure measurements. They found no statistically significant effect of sleeves on the measurement of either systolic or diastolic blood pressure (p > 0.15). However, based on confidence intervals of possible sleeve effects, the authors recommend repeating blood pressure measurements on bared arms when the sleeved-arm oscillometric measurements are at least 86 mm Hg diastolic or 135 mm Hg systolic.

摘要

作者测量了36名分别有裸臂和穿有袖子手臂的受试者的血压,以确定穿袖子对自动示波血压测量的影响。他们发现袖子对收缩压或舒张压的测量没有统计学上的显著影响(p>0.15)。然而,基于袖子可能影响的置信区间,作者建议当穿袖子的手臂进行示波测量时,舒张压至少为86毫米汞柱或收缩压至少为135毫米汞柱时,应在裸臂上重复测量血压。