Suppr超能文献

确保密封植入物外壳中的湿度水平最低。

Ensuring minimal humidity levels in hermetic implant housings.

作者信息

Schuettler Martin, Schatz Andreas, Ordonez Juan S, Stieglitz Thomas

机构信息

Laboratory for Biomedical Microtechnology, Dept of Microsystems Engineering - IMTEK, University of Freiburg, Germany.

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2011;2011:2296-9. doi: 10.1109/IEMBS.2011.6090578.

Abstract

The electronic circuitry of active implantable devices is commonly protected against the risk of water-induced corrosion by using gas-tight (hermetic) packages, preventing moisture from the host body to reach the electronics. However, when closing the package, one has to ensure that the packaged components do not contain moisture that could rise humidity inside the package to critical levels by outgassing. For our miniature metal/ceramic packages, we found a drying procedure of 120 °C at 180 mbar absolute pressure for one hour, followed by a dry helium purge sufficient to keep the relative humidity below 2.5% over a time span of 300 days at 80 °C, corresponding to over 15 years at 37 °C. The additional integration of a desiccant inside the package permits to keep the relative humidity below 0.1%, the detection limit of the integrated sensor. This sensor was selected based on an evaluation of 17 commercially available humidity sensors.

摘要

有源植入式设备的电子电路通常采用气密(密封)封装来防止水致腐蚀的风险,从而防止来自主体的湿气到达电子器件。然而,在封装时,必须确保封装的组件不含湿气,这些湿气可能会因脱气而使封装内部的湿度上升到临界水平。对于我们的微型金属/陶瓷封装,我们发现一种干燥程序:在绝对压力180毫巴下于120°C干燥一小时,随后进行干燥氦气吹扫,足以在80°C下300天的时间内将相对湿度保持在2.5%以下,相当于在37°C下超过15年。在封装内部额外集成干燥剂可使相对湿度保持在0.1%以下,这是集成传感器的检测极限。该传感器是基于对17种市售湿度传感器的评估而选定的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验