Key Laboratory of Analytical Chemistry for Biology and Medicine (Ministry of Education), Department of Chemistry, Wuhan University, Wuhan 430072, China.
Talanta. 2012 Jan 30;89:335-41. doi: 10.1016/j.talanta.2011.12.038. Epub 2011 Dec 20.
Benzimidazole drugs (BZDs) comprise a large number of synthetic anthelmintics, which are widely used in food-producing animals for prophylactic and therapeutic purposes. To protect consumers from the risks related to BZDs residues, a simple, rapid, and efficient method for simultaneous determination of ten BZDs in animal tissues samples was developed. This analytical procedure involved extracting samples with magnetic solid-phase extraction (MSPE) using magnetite/silica/poly (methacrylic acid-co-ethylene glycol dimethacrylate) (Fe(3)O(4)/SiO(2)/poly (MAA-co-EGDMA)) magnetic microspheres, and determination by capillary zone electrophoresis (CZE). To improve the sensitivity of the method, we employed the electrokinetic injection with field-amplified sample stacking technique (FASS). Berbine solution was used as internal standard to minimize the fluctuation of analytical results. Under the optimized extraction conditions, good linearities were obtained for the ten BZDs with the correlation coefficients (R(2)) above 0.9920. The limits of detections (LODs) for ten BZDs were 1.05-10.42 ng/g in swine muscle and 1.06-12.61 ng/g in swine liver, respectively. The intra- and inter-day relative standard deviations (RSDs) of the developed method were less than 13.6%. The recoveries of the ten BZDs for the spiked samples ranged from 81.1% to 105.4% with RSDs less than 9.3%.
苯并咪唑类药物(BZDs)包括大量的合成驱虫药,广泛用于生产食品的动物中,以进行预防和治疗。为了保护消费者免受与 BZDs 残留相关的风险,开发了一种简单、快速、有效的方法,用于同时测定动物组织样品中的十种 BZDs。该分析程序涉及使用磁铁矿/二氧化硅/聚(甲基丙烯酸-co-乙二醇二甲基丙烯酸酯)(Fe(3)O(4)/SiO(2)/poly (MAA-co-EGDMA))磁性微球通过磁性固相萃取(MSPE)提取样品,并通过毛细管区带电泳(CZE)进行测定。为了提高方法的灵敏度,我们采用了场放大样品堆积技术(FASS)的电动进样。伯比林溶液被用作内标,以最小化分析结果的波动。在优化的提取条件下,十种 BZDs 均具有良好的线性关系,相关系数(R(2))均大于 0.9920。十种 BZDs 在猪肌肉中的检出限(LODs)为 1.05-10.42ng/g,在猪肝脏中的检出限(LODs)为 1.06-12.61ng/g。该方法的日内和日间相对标准偏差(RSDs)均小于 13.6%。十种 BZDs 的加标样品回收率在 81.1%至 105.4%之间,RSDs 小于 9.3%。