Unit of Academic Radiology, University of Sheffield, Sheffield, United Kingdom.
Magn Reson Med. 2012 Dec;68(6):1900-4. doi: 10.1002/mrm.24190. Epub 2012 Jan 31.
In this study, the signal-to-noise ratio of hyperpolarized (129)Xe human lung magnetic resonance imaging was compared at 1.5 T and 3 T. Experiments were performed at both B(0) fields with quadrature double Helmholtz transmit-receive chest coils of the same geometry with the same subject loads. Differences in sensitivity between the two field strengths were assessed from the signal-to-noise ratio of multi-slice 2D (129)Xe ventilation lung images obtained at the two field strengths with a spatial resolution of 15 mm × 4 mm × 4 mm. There was a systematically higher signal-to-noise ratio observed at 3 T than at 1.5 T by a factor of 1.25. Mean image signal-to-noise ratio was in the range 27-44 at 1.5 T and 36-51 at 3 T. T 2* of (129)Xe gas in the partially inflated lungs was measured to be 25 ms and 18 ms at 1.5 T and 3 T, respectively. T 2* of (129)Xe gas in fully inflated lungs was measured to be 52 ms and 24 ms at 1.5 T and 3 T, respectively.
在这项研究中,比较了 1.5T 和 3T 下极化(129)Xe 人体肺部磁共振成像的信噪比。在两种 B0 场下进行了实验,使用相同几何形状的正交双亥姆霍兹发射-接收胸部线圈,并对相同的受试者进行了负载。从在两种场强下获得的具有 15mm×4mm×4mm 空间分辨率的多切片 2D(129)Xe 通气肺部图像的信噪比评估了两种场强之间的灵敏度差异。在 3T 下观察到的信噪比比在 1.5T 下系统地高 1.25 倍。在 1.5T 时平均图像信噪比在 27-44 之间,在 3T 时在 36-51 之间。部分充气肺部中(129)Xe 气体的 T2分别为 25ms 和 18ms,在 1.5T 和 3T 下。完全充气肺部中(129)Xe 气体的 T2分别为 52ms 和 24ms,在 1.5T 和 3T 下。