Department of Radiology, UT Health Science Center at San Antonio, TX 78229-3900, USA.
AJR Am J Roentgenol. 2012 Apr;198(4):W347-51. doi: 10.2214/AJR.11.7958.
The purpose of this article is to discuss radiation dose during CT-guided interventions and to explain how radiologists can modify technical factors to minimize radiation doses. Scanner-displayed indexes of radiation exposure that are available during the procedure will be defined to increase awareness about CT radiation dose reduction during interventional procedures.
CT-guided fluoroscopic procedures are safe and effective methods of directed intervention; however, the increasing use of medical radiation is an important consideration. The appropriate use of imaging with an acceptable risk must be considered in every case. During CT-guided interventions, scanner parameters that can be used as a guide for effective dose management, including the CT dose index and dose-length product, are readily displayed. These parameters can be adjusted by modifying the longitudinal scan length, number of scans, and tube current-exposure time product (milliampere × second [mAs]). A team approach to radiation dose reduction will work the best.
本文旨在讨论 CT 引导介入操作中的辐射剂量,并解释放射科医生如何通过调整技术参数来将辐射剂量最小化。本文将定义在介入过程中可用的扫描器显示的辐射暴露指标,以提高对介入过程中 CT 辐射剂量降低的认识。
CT 引导透视操作是安全有效的靶向介入方法;然而,越来越多的医疗辐射的使用是一个重要的考虑因素。在每种情况下,都必须考虑以可接受的风险使用合适的成像方法。在 CT 引导介入中,可用于有效剂量管理的扫描器参数(包括 CT 剂量指数和剂量长度乘积)会被清晰显示。可以通过调整纵向扫描长度、扫描次数以及管电流-曝光时间乘积(毫安×秒[mAs])来调整这些参数。团队合作是降低辐射剂量的最佳方法。