不同消毒剂对眼假肢用丙烯酸树脂的显微硬度和粗糙度的影响。

Effect of different disinfectants on the microhardness and roughness of acrylic resins for ocular prosthesis.

机构信息

Department of Dental Materials and Prosthodontics, UNESP - Univ Estadual Paulista, Faculty of Dentistry of Araçatuba, University of the State of São Paulo, São Paulo, Brazil.

出版信息

Gerodontology. 2013 Mar;30(1):32-9. doi: 10.1111/j.1741-2358.2012.00642.x. Epub 2012 Apr 3.

Abstract

BACKGROUND

Ocular prosthesis materials should have specific properties for their indication and durability; therefore, it is important to investigate their physical behaviour when affected by several disinfectants.

OBJECTIVES

This study evaluated the influence of different disinfecting solutions on the microhardness and surface roughness of acrylic resins for ocular prosthesis.

MATERIALS AND METHODS

Fifty samples simulating ocular prostheses were fabricated with N1 resin and colourless resin and divided (n = 10) according to the disinfectant used: neutral soap, Opti-free, Efferdent, 1% hypochlorite (HYC) and 4% chlorhexidine (CHX). Samples were stored in saline solution at 37°C and disinfected during 120 days. Both microhardness and roughness were investigated before, after 60 days and 120 days of disinfection and storage. Microhardness was measured using a microhardner and the roughness with a roughness device.

RESULTS

N1 resin showed lower microhardness when compared with colourless resin (p < 0.05). HYC and CHX groups exhibited the highest change of microhardness and roughness values (p < 0.05). An increase in roughness and reduction in microhardness of ocular acrylic resins were observed after both periods of disinfection and storage (p < 0.05).

CONCLUSION

Both disinfection/storage periods affected the microhardness and roughness values of the samples.

摘要

背景

眼部假肢材料应具有特定的性能,以满足其指示和耐用性的要求;因此,研究它们在受到几种消毒剂影响时的物理性能是很重要的。

目的

本研究评估了不同消毒剂对眼部假肢用丙烯酸树脂的显微硬度和表面粗糙度的影响。

材料和方法

用 N1 树脂和无色树脂制作了 50 个模拟眼部假肢的样本,并根据使用的消毒剂(中性肥皂、Opti-free、Efferdent、1%次氯酸钠[HYC]和 4%氯己定[CHX])将其分为(n = 10)。样本在 37°C 的盐溶液中储存,并在 120 天内进行消毒。在消毒和储存前、60 天和 120 天后,分别对显微硬度和粗糙度进行了检测。使用显微硬度计测量显微硬度,使用粗糙度仪测量粗糙度。

结果

N1 树脂的显微硬度明显低于无色树脂(p < 0.05)。HYC 和 CHX 组的显微硬度和粗糙度值变化最大(p < 0.05)。在消毒和储存两个时期后,眼部丙烯酸树脂的粗糙度增加,显微硬度降低(p < 0.05)。

结论

消毒/储存两个时期都影响了样本的显微硬度和粗糙度值。