Department of Anesthesiology, National Medical Center, Seoul, Korea.
Korean J Anesthesiol. 2012 Mar;62(3):234-9. doi: 10.4097/kjae.2012.62.3.234. Epub 2012 Mar 21.
For patients in the intensive care unit (ICU) or under monitored anesthetic care (MAC), the precise monitoring of sedation depth facilitates the optimization of dosage and prevents adverse complications from underor over-sedation. For this purpose, conventional subjective sedation scales, such as the Observer's Assessment of Alertness/Sedation (OAA/S) or the Ramsay scale, have been widely utilized. Current procedures frequently disturb the patient's comfort and compromise the already well-established sedation. Therefore, reliable objective sedation scales that do not cause disturbances would be beneficial. We aimed to determine whether spectral entropy can be used as a sedation monitor as well as determine its ability to discriminate all levels of propofol-induced sedation during gradual increments of propofol dosage.
In 25 healthy volunteers undergoing general anesthesia, the values of response entropy (RE) and state entropy (SE) corresponding to each OAA/S (5 to 1) were determined. The scores were then analyzed during each 0.5 mcg/ml- incremental increase of a propofol dose.
We observed a reduction of both RE and SE values that correlated with the OAA/S (correlation coefficient of 0.819 in RE-OAA/S and 0.753 in SE-OAA/S). The RE and SE values corresponding to awake (OAA/S score 5), light sedation (OAA/S 3-4) and deep sedation (OAA/S 1-2) displayed differences (P < 0.05).
The results indicate that spectral entropy can be utilized as a reliable objective monitor to determine the depth of propofol-induced sedation.
对于重症监护病房(ICU)中的患者或接受监测麻醉护理(MAC)的患者,精确监测镇静深度有助于优化剂量,并防止镇静不足或过度引起的不良并发症。为此,传统的主观镇静量表,如观察者评估警觉/镇静量表(OAA/S)或 Ramsay 量表,已被广泛应用。目前的操作程序经常会干扰患者的舒适度,并影响已经确立的镇静效果。因此,使用不干扰患者的可靠客观镇静量表将是有益的。我们旨在确定光谱熵是否可作为镇静监测指标,以及其在逐步增加异丙酚剂量时对所有异丙酚诱导镇静水平的区分能力。
在 25 名接受全身麻醉的健康志愿者中,确定与每个 OAA/S(5 至 1)相对应的反应熵(RE)和状态熵(SE)的值。然后在每个 0.5 mcg/ml-异丙酚剂量递增时分析这些评分。
我们观察到 RE 和 SE 值的降低与 OAA/S 相关(RE-OAA/S 的相关系数为 0.819,SE-OAA/S 的相关系数为 0.753)。对应于清醒(OAA/S 评分 5)、轻度镇静(OAA/S 3-4)和深度镇静(OAA/S 1-2)的 RE 和 SE 值存在差异(P<0.05)。
结果表明,光谱熵可用作可靠的客观监测指标,以确定异丙酚诱导镇静的深度。