Suppr超能文献

采用气相色谱-火焰离子化检测法同时测定大鼠尿液中苯乙烯和氧化苯乙烯的芳香酸代谢物。

Simultaneous determination of aromatic acid metabolites of styrene and styrene-oxide in rat urine by gas chromatography-flame ionization detection.

机构信息

Institut National de Recherche et de Sécurité (INRS), Rue du Morvan CS 60027, Vandoeuvre, 54519 cedex, France.

出版信息

J Anal Toxicol. 2012 Jun;36(5):312-8. doi: 10.1093/jat/bks015. Epub 2012 Apr 5.

Abstract

A convenient and reliable gas chromatographic method was developed for the simultaneous determination of six aromatic acid metabolites of styrene and styrene-oxide in rat urine; i.e., benzoic (BA), phenylacetic (PAA), mandelic (MA), phenylglyoxylic (PGA), hippuric (HA) and phenylaceturic (PAUA) acids. The method involves a one-pot esterification-extraction procedure, performed directly on urine without prior treatment. Analyses were performed on a RTX-1701 capillary column and the recovered isopropyl esters derivatives were detected by flame ionization detection. The analytical method was validated for selectivity, linearity, detection and quantification limits, recovery and intra-day and inter-day precisions. Calibration curves showed linearity in the range of 8-800 mg/L, except for HA and PAUA (40-800 mg/L). Limits of detection were between 0.2 (PPA) and 7.0 (PAUA) mg/L. The intra-day precisions determined at three concentrations levels were less than 5% for BA, PAA, MA and PGA and 9% for HA and PAUA, respectively. The corresponding mean inter-day precisions for these two groups were 8 and 16%, respectively. The method was successfully applied to quantitatively analyze styrene, styrene-oxide, ethylbenzene and toluene metabolites in urine samples from rats exposed by inhalation to these compounds at levels close to the occupational threshold limit values. Provided that this method can be transposed to human urine, it could have applications as part of biological monitoring for workers exposed to styrene or related compounds.

摘要

建立了一种同时测定大鼠尿液中苯乙烯和氧化苯乙烯 6 种芳香酸代谢物的简便、可靠的气相色谱法;即苯甲酸(BA)、苯乙酸(PAA)、扁桃酸(MA)、苯乙醛酸(PGA)、马尿酸(HA)和苯乙酰脲酸(PAUA)。该方法涉及一锅酯化-萃取程序,可直接在尿液中进行,无需预先处理。分析在 RTX-1701 毛细管柱上进行,并用火焰电离检测检测回收的异丙基酯衍生物。该分析方法经过了选择性、线性、检测限和定量限、回收率和日内及日间精密度的验证。校准曲线在线性范围内(8-800mg/L),除 HA 和 PAUA 外(40-800mg/L)。检测限在 0.2(PPA)至 7.0(PAUA)mg/L 之间。在三个浓度水平下测定的日内精密度,BA、PAA、MA 和 PGA 的精密度小于 5%,HA 和 PAUA 的精密度分别为 9%。这两组的相应平均日间精密度分别为 8%和 16%。该方法成功应用于定量分析大鼠尿液中苯乙烯、氧化苯乙烯、乙苯和甲苯代谢物,这些化合物的吸入暴露水平接近职业阈限值。假设该方法可以转移到人体尿液中,它可以作为接触苯乙烯或相关化合物的工人生物监测的一部分。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验