Cardiology Service, University Hospital, 4, rue Gabrielle-Perret-Gentil, 1211 Geneva 14, Switzerland.
Eur Heart J Cardiovasc Imaging. 2013 Jan;14(1):38-42. doi: 10.1093/ehjci/jes085. Epub 2012 Apr 25.
Pocket-size echographs may be useful for bedside diagnosis in acute cardiac care, but their diagnostic accuracy in this setting has not been well tested. Our aim was to evaluate this tool in patients requiring an urgent echocardiogram.
Trained cardiologists performed echocardiograms with a pocket-size echograph (Vscan) in consecutive patients requiring urgent echocardiography. The exams were then compared in a blinded manner with echocardiograms performed with a high-end standard echocardiograph.
A total of 104 patients were studied. There was an excellent agreement between the Vscan and the high-end echocardiograph for the left ventricular systolic function and pericardial effusion (Kappa: 0.89 and 0.81, respectively), and the agreement was good or moderate for evaluating the aortic, mitral, and tricuspid valve function and the left ventricular size (Kappa: 0.55-0.66). Visualization of the Vscan images in full-screen format on a PC did not in general confer added value.
The Vscan used by a trained cardiologist has good diagnostic accuracy in the emergency setting compared with a high-end echocardiograph, despite small screen size and lack of pulse-wave and continuous Doppler.
口袋大小的超声心动图仪可能有助于在急性心脏护理中进行床边诊断,但尚未对其在该环境下的诊断准确性进行充分测试。我们的目的是评估该工具在需要紧急超声心动图检查的患者中的应用。
经过培训的心脏病专家使用口袋大小的超声心动图仪(Vscan)对连续需要紧急超声心动图检查的患者进行超声心动图检查。然后,将这些检查与使用高端标准超声心动图仪进行的超声心动图检查进行盲法比较。
共研究了 104 例患者。Vscan 与高端超声心动图仪在左心室收缩功能和心包积液方面具有极好的一致性(Kappa:0.89 和 0.81),而在评估主动脉瓣、二尖瓣和三尖瓣功能以及左心室大小方面的一致性为良好或中度(Kappa:0.55-0.66)。在 PC 上全屏显示 Vscan 图像通常不会增加价值。
尽管屏幕较小且缺乏脉冲波和连续多普勒,但经过训练的心脏病专家使用的 Vscan 与高端超声心动图仪相比,在紧急情况下具有良好的诊断准确性。