Ouyang Ping-Bo, Li Cong-Yi, Zhu Xiao-Hua, Duan Xuan-Chu
Department of Ophthalmology, the Second Xiangya Hospital of Central South University, Changsha 410011, Hunan Province, China.
Int J Ophthalmol. 2012;5(1):102-7. doi: 10.3980/j.issn.2222-3959.2012.01.21. Epub 2012 Feb 18.
To investigate the accuracy of intraocular pressure (IOP) as measured by a Reichert Ocular Response Analyzer (ORA), as well as the relationship between central corneal thickness (CCT) and IOP as measured by ORA, Goldmann applanation tonometry (GAT), and dynamic contour tonometry (DCT).
A total of 158 healthy individuals (296 eyes) were chosen randomly for measurement of IOP. After CCT was measured using A-ultrasound (A-US), IOP was measured by ORA, GAT, and DCT devices in a randomized order. The IOP values acquired using each of the three tonometries were compared, and the relationship between CCT and IOP values were analyzed separately. Two IOP values, Goldmann-correlated IOP value (IOPg) and corneal-compensated intraocular pressure (IOPcc), were got using ORA. Three groups were defined according to CCT: 1) thin cornea (CCT<520µm); 2) normal-thickness cornea (CCT: 520-580µm); and 3) thick cornea (CCT>580µm) groups.
In normal subjects, IOP measurements were 14.95±2.99mmHg with ORA (IOPg), 15.21±2.77mmHg with ORA (IOPcc), 15.22±2.77mmHg with GAT, and 15.49±2.56mmHg with DCT. Mean differences were 0.01±2.29mmHg between IOPcc and GAT (P>0.05) and 0.28±2.20mmHg between IOPcc and DCT (P>0.05). There was a greater correlation between IOPcc and DCT (r=0.946, P=0.000) than that between IOPcc and GAT (r=0.845, P=0.000). DCT had a significant correlation with GAT (r=0.854, P=0.000). GAT was moderately correlated with CCT (r=0.296, P<0.001), while IOPcc showed a weak but significant correlation with CCT (r=-0.155, P=0.007). There was a strong negative correlation between CCT and the difference between IOPcc and GAT(r=-0.803, P=0.000), with every 10µm increase in CCT resulting in an increase in this difference of 0.35mmHg. The thick cornea group (CCT>580µm) showed the least significant correlation between IOPcc and GAT (r=0.859, P=0.000); while the thin cornea group (CCT<520µm) had the most significant correlation between IOPcc and GAT (r=0.926, P=0.000). The correlated differences between IOPcc and DCT were not significant in any of the three groups (P>0.05).
Measurement of IOP by ORA has high repeatability and is largely consistent with GAT measurements. Moreover, the ORA measurements are affected only to a small extent by CCT, and are likely to be much closer to the real IOP value than GAT.
研究瑞特眼反应分析仪(ORA)测量眼压(IOP)的准确性,以及中央角膜厚度(CCT)与通过ORA、Goldmann压平眼压计(GAT)和动态轮廓眼压计(DCT)测量的IOP之间的关系。
随机选取158名健康个体(296只眼)测量IOP。使用A超(A-US)测量CCT后,以随机顺序通过ORA、GAT和DCT设备测量IOP。比较使用三种眼压计获得的IOP值,并分别分析CCT与IOP值之间的关系。使用ORA获得两个IOP值,即Goldmann相关眼压值(IOPg)和角膜补偿眼压(IOPcc)。根据CCT定义三组:1)薄角膜(CCT<520µm);2)正常厚度角膜(CCT:520 - 580µm);3)厚角膜(CCT>580µm)组。
在正常受试者中,ORA测量的IOP(IOPg)为14.95±2.99mmHg,ORA测量的IOP(IOPcc)为15.21±2.77mmHg,GAT测量的IOP为15.22±2.77mmHg,DCT测量的IOP为来15.49±2.56mmHg。IOPcc与GAT之间的平均差异为0.01±2.29mmHg(P>0.05),IOPcc与DCT之间的平均差异为0.28±2.20mmHg(P>0.05)。IOPcc与DCT之间的相关性(r = 0.946,P = 0.000)高于IOPcc与GAT之间的相关性(r = 0.845,P = 0.000)。DCT与GAT有显著相关性(r = 0.8富54,P = 0.000)。GAT与CCT中度相关(r = 0.296,P<0.001),而IOPcc与CCT呈弱但显著的相关性(r = -0.155,P = 0.007)。CCT与IOPcc和GAT之间的差异呈强负相关(r = -0.803,P = 0.000),CCT每增加10µm,该差异增加0.35mmHg。厚角膜组(CCT>富580µm)中IOPcc与GAT之间的相关性最不显著(r = 0.859,P = 0.000);而薄角膜组(CCT<520µm)中IOPcc与GAT之间的相关性最显著(r = 0.926,P = 0.000)。IOPcc与DCT之间的相关差异在三组中均不显著(P>0.05)。
ORA测量IOP具有高重复性,且与GAT测量结果基本一致。此外,ORA测量受CCT影响较小,可能比GAT更接近真实的IOP值。