Intraocular Pressure Measurement Using Ocular Response Analyzer, Dynamic Contour Tonometer, and Scheimpflug Analyzer Corvis ST.

作者信息

Ramm Lisa, Herber Robert, Spoerl Eberhard, Raiskup Frederik, Pillunat Lutz E, Terai Naim

机构信息

Department of Ophthalmology, University Hospital Carl Gustav Carus, TU Dresden Fetscherstraße 74, 01307 Dresden, Germany.

出版信息

J Ophthalmol. 2019 Oct 16;2019:3879651. doi: 10.1155/2019/3879651. eCollection 2019.

Abstract

PURPOSE

To compare intraocular pressure (IOP) measurements with Goldmann applanation tonometry (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST (CST) in healthy subjects.

METHODS

In a prospective, observational study, IOP measurements with GAT (GAT-IOPc), ORA (IOPcc), DCT (DCT-IOP), and CST (bIOP) were performed and analyzed in 94 healthy subjects.

RESULTS

Mean age of the participants was 45.6 ± 17.2 years (range 18 to 81 years). Mean GAT-IOPc was 12.9 ± 2.4 mmHg, mean DCT-IOP was 16.1 ± 2.6 mmHg, and mean IOPcc was 15.6 ± 3.3 mmHg. DCT-IOP and IOPcc were significantly higher than GAT-IOPc ( < 0.001). Mean bIOP was 13.5 ± 2.4 mmHg that was slightly higher but not significantly different from GAT-IOPc (=0.146). Correlation analysis of IOP values and central corneal thickness (CCT) revealed a negative correlation between GAT-IOPc and CCT ( = -0.347; =0.001). However, IOPcc, DCT-IOP, and bIOP showed no significant correlation to CCT. Only bIOP revealed a weak but significant age dependency ( = 0.321, =0.002).

CONCLUSION

All tonometry devices showed a good agreement of biomechanical corrected IOP values with GAT-IOPc. As no influence of CCT on IOPcc, DCT-IOP, and bIOP was detectable, the used correction algorithms appear to be appropriate in these tonometers in the clinical setting. The highest agreement was found between GAT-IOPc and bIOP. However, bIOP weakly correlated with participants' age. Further studies are needed to elucidate the role of bIOP for IOP measurement.

摘要

目的

比较健康受试者使用Goldmann压平眼压计(GAT)、眼反应分析仪(ORA)、动态轮廓眼压计(DCT)和Corvis ST(CST)测量眼压(IOP)的情况。

方法

在一项前瞻性观察研究中,对94名健康受试者进行了GAT(GAT-IOPc)、ORA(IOPcc)、DCT(DCT-IOP)和CST(bIOP)的眼压测量并进行分析。

结果

参与者的平均年龄为45.6±17.2岁(范围18至81岁)。平均GAT-IOPc为12.9±2.4 mmHg,平均DCT-IOP为16.1±2.6 mmHg,平均IOPcc为15.6±3.3 mmHg。DCT-IOP和IOPcc显著高于GAT-IOPc(<0.001)。平均bIOP为13.5±2.4 mmHg,略高于GAT-IOPc但无显著差异(=0.146)。眼压值与中央角膜厚度(CCT)的相关性分析显示,GAT-IOPc与CCT呈负相关(=-0.347;=0.001)。然而,IOPcc、DCT-IOP和bIOP与CCT无显著相关性。仅bIOP显示出微弱但显著的年龄依赖性(=0.321,=0.002)。

结论

所有眼压计的生物力学校正眼压值与GAT-IOPc均显示出良好的一致性。由于未检测到CCT对IOPcc、DCT-IOP和bIOP的影响,在临床环境中,这些眼压计所使用的校正算法似乎是合适的。GAT-IOPc与bIOP之间的一致性最高。然而,bIOP与参与者年龄呈弱相关。需要进一步研究以阐明bIOP在眼压测量中的作用。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/5715/6815996/30a6a27a14af/JOPH2019-3879651.001.jpg